NC TRAVELER CORONARY DILATATION CATHETER
Report
- Report Number
- 2024168-2017-00575
- Event Type
- Malfunction
- Date Received
- January 19, 2017
- Date of Event
- December 26, 2016
- Report Date
- January 31, 2017
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- LOX
- PMA / PMN Number
- K110134
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. CONCOMITANT MEDICAL PRODUCTS: GUIDE WIRE: SION, GUIDE CATHETER: HYPERION, STENT: XIENCE ALPINE 3.0X38MM. THE CUSTOMER REPORTED THE DEVICE WAS DISCARDED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL RELEVANT INFORMATION. THE NC TRAVELER IS CURRENTLY NOT COMMERCIALLY AVAILABLE IN THE U.S.; HOWEVER, IT IS SIMILAR TO A DEVICE SOLD IN THE U.S.
(B)(4). THE DEVICE WAS NOT RETURNED FOR EVALUATION. A REVIEW OF THE LOT HISTORY RECORD IDENTIFIED NO MANUFACTURING NONCONFORMITIES ISSUED TO THE REPORTED LOT THAT WOULD HAVE CONTRIBUTED TO THIS EVENT. ADDITIONALLY, A REVIEW OF THE COMPLAINT HISTORY IDENTIFIED NO OTHER INCIDENTS FROM THIS LOT. THE INVESTIGATION DETERMINED THE REPORTED DIFFICULTIES APPEAR TO BE RELATED TO CIRCUMSTANCES OF THE PROCEDURE AND THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECTS TO THE DESIGN, MANUFACTURE OR LABELING OF THE DEVICE.
IT WAS REPORTED THAT DURING A PROCEDURE OF THE MILDLY TORTUOUS, ECCENTRIC, MODERATELY CALCIFIED, DISTAL RIGHT CORONARY ARTERY (RCA), AFTER STENT IMPLANTATION, THE 3.5 X 20 MM NC TRAVELER BALLOON CATHETER WAS ADVANCED, BUT MET RESISTANCE WITH THE STENT. DURING INFLATION, THE BALLOON RUPTURED AT 10 ATMOSPHERE (ATM). THE BALLOON WAS REMOVED AND A DIFFERENT 3.5 X 15 MM NC TRAVELER BALLOON WAS USED IN THE PROCEDURE. THERE WAS NO REPORTED ADVERSE PATIENT EFFECT. THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 46759 | NC TRAVELER CORONARY DILATATION CATHETER | CORONARY DILATATION CATHETER | LOX | AV-TEMECULA-CT | 50904G1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |