FDA Adverse Event Malfunction Summary report: N

STERRAD® CYCLESURE® 24 BIOLOGICAL INDICATOR

MDR report key: 6263812 · Received January 19, 2017

Report

Report Number
2084725-2017-00031
Event Type
Malfunction
Date Received
January 19, 2017
Date of Event
December 28, 2016
Report Date
December 29, 2016
Manufacturer
ADVANCED STERILIZATION PRODUCTS
Product Code
FRC
PMA / PMN Number
K994055
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DHR, COMPLAINT TRENDING, RETAINS TESTING, CONCOMITANT PRODUCT EVALUATION AND VISUAL ANALYSIS WERE NOT PERFORMED AS THERE IS SUFFICIENT INFORMATION TO DETERMINE USER ERROR AS THE CAUSE OF THE ISSUE. THE SRA INDICATES THE RISK ASSOCIATED WITH A QUALITY PROBLEM WITH NO IMPACT ON SAFETY IS "LOW." THE ASSIGNABLE CAUSE OF THE ISSUE CAN BE ATTRIBUTED TO USER ERROR AS THE CHEMICAL INDICATOR (CI) STRIPS WERE PLACED INCORRECTLY IN THE CONCOMITANT STERRAD® 100NX UNIT. THE CUSTOMER STATED THE CI STRIPS WERE BEING STORED INSIDE A POUCH UNDERNEATH A LOADED TRAY. THE TSR PROVIDED INSTRUCTIONS OVER THE PHONE TO THE CUSTOMER AND A LETTER WAS SENT WITH THE INSTRUCTIONS FOR USE (IFU) IN REGARDS TO PROPER LOAD PLACEMENT. THIS ISSUE WILL CONTINUE TO BE TRACKED AND TRENDED.

Additional Manufacturer Narrative · 1

THE CORRECT BRAND NAME IS STERRAD® CHEMICAL INDICATOR STRIP. THE CORRECT COMMON DEVICE IS INDICATOR, CHEMICAL. THE CORRECT CATALOG NUMBER IS 14100_90. THE CORRECT LOT NUMBER IS 132611-03.

Description of Event or Problem · 1

A CUSTOMER REPORTED A STERRAD® CHEMICAL INDICATOR STRIP DID NOT CHANGE COLOR CORRECTLY AFTER A COMPLETED EXPRESS CYCLE ON THE STERRAD® 100NX CYCLE. THE AFFECTED LOAD WAS NOT RELEASED. THERE WAS NO REPORT OF INFECTION, INJURY OR HARM TO PATIENT(S) ASSOCIATED WITH THIS ISSUE. ALTHOUGH THERE IS NO REPORT OF PATIENT INJURY OR HARM AND NO PRIOR INCIDENTS HAVE RESULTED IN SERIOUS INJURY, ADVANCED STERILIZATION PRODUCTS (ASP) HAS DETERMINED IN THIS SITUATION STERILITY CANNOT BE ASSURED. THEREFORE, AS A MATTER OF POLICY ASP HAD DECIDED TO REPORT ALL INCIDENTS OF STERRAD® CHEMICAL INDICATOR STRIPS NOT CHANGING COLOR CORRECTLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
47229 STERRAD® CYCLESURE® 24 BIOLOGICAL INDICATOR INDICATOR, BIOLOGICAL (FRC) FRC ADVANCED STERILIZATION PRODUCTS NA

Patients

Seq Age Sex Outcome Treatment
1