FDA Adverse Event Malfunction Summary report: N

TM-400

MDR report key: 6261009 · Received January 18, 2017

Report

Report Number
9614750-2017-00003
Event Type
Malfunction
Date Received
January 18, 2017
Date of Event
March 14, 2016
Report Date
January 18, 2017
Manufacturer
ITO CO., LTD. TSUKUBA FACTORY
Product Code
ITH
PMA / PMN Number
K081247
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE SUBJECT DEVICE WAS DISCARDED AT THE EVENT SITE. DEVICE NOT RETURNED TO MANUFACTURER.

Additional Manufacturer Narrative · 1

ITO EXAMINED THE DEVICE HISTORY RECORD FOR THE SUBJECT TM-400 DEVICE. THERE WERE NO PROBLEMS OBSERVED DURING THE MANUFACTURING OR TESTING NOTED IN THE DHR. THE USER FACILITY DID NOT DISCLOSE THE PATIENT'S INFORMATION.

Description of Event or Problem · 1

DURING A PROCEDURE, THE TRACTION ROPE OF THE SUBJECT TM-400 DEVICE GOT STUCK. EVEN AFTER WAITING A WHILE TO TRY TO PULL OUT THE ROPE, THE ROPE WAS STILL STUCK AND COULD NOT BE PULLED OUT. AFTER TURNING OFF THE POWER AND TRYING TO RESUME THE PROCEDURE, THE ROPE COULD BE REELED UP AND A SMOOTH TRACTION WAS RESUMED. WHEN THE TRACTION POWER LEVEL WAS SET FROM 0 TO 60%, THE ROPE GOT STUCK INSTANTANEOUSLY AND THEN STARTED TRACTION WITH A HIGH SPEED (A SPEED OF 60 TO 100% TRACTION POWER). WHEN THE 90KG SETTING WAS SELECTED, THE ROPE BECAME STUCK WITHOUT EXCEPTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
43399 TM-400 EQUIPMENT, TRACTION, POWERED ITH ITO CO., LTD. TSUKUBA FACTORY TM-400

Patients

Seq Age Sex Outcome Treatment
1 Other