TM-400
Report
- Report Number
- 9614750-2017-00003
- Event Type
- Malfunction
- Date Received
- January 18, 2017
- Date of Event
- March 14, 2016
- Report Date
- January 18, 2017
- Manufacturer
- ITO CO., LTD. TSUKUBA FACTORY
- Product Code
- ITH
- PMA / PMN Number
- K081247
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TW
- Reporter Occupation
- OTHER
Narratives
THE SUBJECT DEVICE WAS DISCARDED AT THE EVENT SITE. DEVICE NOT RETURNED TO MANUFACTURER.
ITO EXAMINED THE DEVICE HISTORY RECORD FOR THE SUBJECT TM-400 DEVICE. THERE WERE NO PROBLEMS OBSERVED DURING THE MANUFACTURING OR TESTING NOTED IN THE DHR. THE USER FACILITY DID NOT DISCLOSE THE PATIENT'S INFORMATION.
DURING A PROCEDURE, THE TRACTION ROPE OF THE SUBJECT TM-400 DEVICE GOT STUCK. EVEN AFTER WAITING A WHILE TO TRY TO PULL OUT THE ROPE, THE ROPE WAS STILL STUCK AND COULD NOT BE PULLED OUT. AFTER TURNING OFF THE POWER AND TRYING TO RESUME THE PROCEDURE, THE ROPE COULD BE REELED UP AND A SMOOTH TRACTION WAS RESUMED. WHEN THE TRACTION POWER LEVEL WAS SET FROM 0 TO 60%, THE ROPE GOT STUCK INSTANTANEOUSLY AND THEN STARTED TRACTION WITH A HIGH SPEED (A SPEED OF 60 TO 100% TRACTION POWER). WHEN THE 90KG SETTING WAS SELECTED, THE ROPE BECAME STUCK WITHOUT EXCEPTION.