FDA Adverse Event Malfunction Summary report: N

PRECISION XTRA

MDR report key: 6254995 · Received January 16, 2017

Report

Report Number
2954323-2017-00390
Event Type
Malfunction
Date Received
January 16, 2017
Date of Event
January 4, 2017
Report Date
January 29, 2018
Manufacturer
ABBOTT DIABETES CARE INC
Product Code
NBW
PMA / PMN Number
K040814
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THIS SERVES AS A CORRECTION REPORT. BRAND NAME WAS INCORRECTLY IDENTIFIED AS FREESTYLE PRECISION NEO. THE CORRECT BRAND NAME FOR THE REPORTED METER SERIAL NUMBER IS (B)(4). BRAND NAME HAS BEEN UPDATED TO REFLECT THE CORRECTION. IN ADDITION PMA/510K# HAS BEEN UPDATED TO K040814, AND CATALOG NO HAS BEEN UPDATED TO 98814-65.

Additional Manufacturer Narrative · 1

AN EXTENDED INVESTIGATION HAS BEEN CONDUCTED, WHICH INVOLVED A MANUFACTURING REVIEW (INCLUDING 5 YEARS OF DEVICE HISTORY RECORDS (DHRS)), PREVIOUS COMPLAINT AND CORRECTIVE AND PREVENTIVE ACTION (CAPA) INVESTIGATIONS CONDUCTED FOR BLANK SCREEN ISSUES, PROCESS FAILURE MODE EFFECTS ANALYSES (PFMEAS), DESIGN CONTROLS, AND DESIGN FAILURE MODE EFFECTS ANALYSES (DFMEAS), RISK MANAGEMENT REPORTS, RISK EVALUATIONS, AND LABEL COPY. THE INVESTIGATION DID NOT IDENTIFY ANY INDICATION THAT THE PRODUCT DID NOT MEET SPECIFICATION. IF THE PRODUCT IS RETURNED, AN INVESTIGATION WILL BE PERFORMED.

Additional Manufacturer Narrative · 1

UNLESS FURTHER INFORMATION IS OBTAINED, THIS ISSUE IS CONSIDERED CLOSED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.

Description of Event or Problem · 1

THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Description of Event or Problem · 1

THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.