FDA Adverse Event Malfunction Summary report: N

MODULAR EBS KNEE JOINT

MDR report key: 6253227 · Received January 16, 2017

Report

Report Number
9616494-2017-00003
Event Type
Malfunction
Date Received
January 16, 2017
Report Date
October 5, 2016
Manufacturer
OTTO BOCK HEALTHCARE GMBH
Product Code
ISY
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE COMPLAINT NOTIFICATION ASSOCIATED WITH THIS DEVICE DESCRIBED THAT THE BUSHINGS ARE BROKEN CAUSING THE BOLTS TO BACK OUT. NO SERIOUS INJURIES WERE SUSTAINED AS A RESULT OF THE FAILURE AND NO MEDICAL TREATMENT WAS REQUIRED. THE EVALUATION OF THIS DEVICE WAS CONDUCTED AT THE MANUFACTURER'S SERVICE CENTER ON NOVEMBER 21ST, 2016. AFTER EVALUATION, WE CAN CONFIRM THAT ALL BOLTS HAD LOOSENED. WE CAN EXCLUDE A MANUFACTURING FAULT. FURTHER DATA REGARDING USAGE AND PATIENT WERE REQUESTED. AFTER TWO ATTEMPTS TO GET THESE DATA, THEY ARRIVED DECEMBER 19TH, 2016. THE FINAL EVALUATION WAS CONDUCTED ON JANUARY 6TH, 2017. AN EXACT REASON FOR THE LOOSE BOLTS COULD NOT BE DETERMINED. NO FALL OR INJURY OCCURRED DUE TO THIS FAILURE, BUT IT COULD HAVE LED TO PATIENT INJURY.

Description of Event or Problem · 1

KNEE JOINT WAS SENT IN FOR REPAIR. CUSTOMER STATED THAT THE BUSHINGS ARE BROKEN CAUSING THE BOLTS TO BACK OUT. NO FALL OR INJURY OCCURRED DUE TO THIS FAILURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
36235 MODULAR EBS KNEE JOINT MODULAR EBS KNEE JOINT ISY OTTO BOCK HEALTHCARE GMBH 3R60

Patients

Seq Age Sex Outcome Treatment
1