NAKOMA-SL ACP SYSTEM
Report
- Report Number
- 3008520015-2016-00002
- Event Type
- Malfunction
- Date Received
- January 12, 2017
- Date of Event
- September 9, 2016
- Report Date
- October 7, 2016
- Manufacturer
- ALLIANCE PARTNERS LLC (DBA ALLIANCE SPINE)
- Product Code
- KWQ
- PMA / PMN Number
- K141993
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
THE NAKOMA-SL 1-LEVEL PLATE INVOLVED IN THE EVENT, PART # 3100I-1005 LOT # 21672S, WAS RECEIVED FROM THE CONTRACTED MANUFACTURER AS PART OF A LOT QUANTITY OF (B)(4) ON 11/2/2014. THE LOT WAS INSPECTED AND RELEASED TO INVENTORY. AN ENGINEERING EVALUATION WAS COMPLETED ON THE DEVICES INVOLVED IN THE REPORTED EVENT. EVALUATION INCLUDED VISUAL AND DIMENSIONAL INSPECTION OF THE RETURNED 22 MM 1-LEVEL PLATE (31001-1005; LOT #21672S), FOUR (4) SELF-TAPPING SCREWS (PART #3100I-6004; LOT #24347S) AND FIXED SINGLE BARREL DRILL GUIDE (PART #3100T-1400; LOT # SM60277), AND THEN SIMULATED USE TESTING IN AN ATTEMPT TO RECREATE THE REPORTED EVENT USING THE RETURNED DEVICES BY IMPLANTING EACH OF THE FOUR (4) SCREWS INTO EACH OF THE PLATE'S FOUR (4) SCREW HOLES PER THE SURGICAL TECHNIQUE USING THE RETURNED DRILL GUIDE, A DRILL BIT THAT WAS THE SAME CONFIGURATION AS THE INSTRUMENT REFERENCED IN THE REPORTED EVENT, AND A SAW BONE. RESULTS WERE AS FOLLOWS: DEVICES WERE WITHIN SPECIFICATION; AND ZERO FAILURES WERE OBSERVED DURING THE SIMULATED USE EVALUATION; ALL FOUR (4) SCREWS WERE SUCCESSFULLY IMPLANTED IN EACH OF THE FOUR (4) PLATE HOLES WITHOUT PASSING THRU THE PLATE. OBSERVED THE MATING CONNECTION BETWEEN THE DRILL GUIDE AND PLATE HOLE WAS SLIGHTLY UNSTABLE. BASED ON THE EVALUATION PERFORMED, THE FAILURE MODE COULD NOT BE RE-CREATED (NOT CONFIRMED). PARTS INVOLVED IN THE REPORTED EVENT WERE FOUND TO BE WITHIN SPECIFICATION WITH NO IRREGULARITIES IDENTIFIED AND TESTING COULD NOT SUBSTANTIATE THE SCREW PASSING THRU THE PLATE FAILURE MODE ALLEGED IN THE COMPLAINT. IT WAS OBSERVED, HOWEVER, THAT THE MATING CONNECTION BETWEEN THE DRILL GUIDE AND PLATE HOLE WAS SLIGHTLY UNSTABLE, WHICH COULD AFFECT PROPER SCREW HOLE PREPARATION. UNSTABLE CONNECTION AT THIS INTERFACE COULD CAUSE THE DRILLED HOLE TO BE OFF AXIS, RESULTING IN A DRILLED HOLE OUTSIDE THE NAKOMA-SL SYSTEM'S ALLOWABLE 7-DEGREE ANGULATION.
ON (B)(6) 2016, THE DOCTOR WAS PERFORMING A C5-C7 ACDF PROCEDURE. DURING IMPLANTATION OF A NAKOMA-SL 22 MM 1-LEVEL PLATE (PART 3100I-1005 LOT #21672S) AND FOUR (4) NAKOMA-SL 4.0 MM X 14 MM SELF-TAPPING SCREWS (PART #3100I-6004 LOT #24347S), THE SCREWS PASSED THROUGH THE PLATE. A FIXED SINGLE BARREL DRILL GUIDE (PART #3100T-1400 LOT #SM60277) AND A 2.3 MM X 12 MM DRILL BIT WAS BEING USED TO PREPARE THE HOLES DURING THE SURGERY. ALL FOUR (4) SCREWS AND PLATE WERE SAFELY REMOVED FROM THE PATIENT AND REPLACED WITH ANOTHER SYSTEM. NO PATIENT INJURIES WERE REPORTED. THE NAKOMA-SL 1-LEVEL PLATE (31001-1005; LOT #21672S), AND THE FOUR (4) SELF-TAPPING SCREWS (PART #3100I-6004; LOT #24347S) WERE RETURNED FOR PRODUCT EVALUATION. IN ADDITION, THE FIXED SINGLE BARREL DRILL GUIDE THAT WAS USED DURING THE PROCEDURE (PART # 3100T-1400; LOT # SM60277) WAS ALSO RETURNED. THIS EVENT INVOLVED TWO DEVICES. THIS IS REPORT 1 OF 2.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 28760 | NAKOMA-SL ACP SYSTEM | ANTERIOR CERVICAL PLATING SYSTEM | KWQ | ALLIANCE PARTNERS LLC (DBA ALLIANCE SPINE) | 3100I-1005 | 21672S |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 49 YR |