CORTEX SCREW AXSOS 3 TI Ø3.5MM / L36MM
Report
- Report Number
- 0008031020-2017-00025
- Event Type
- Injury
- Date Received
- January 12, 2017
- Date of Event
- October 30, 2016
- Report Date
- March 10, 2017
- Manufacturer
- STRYKER GMBH
- Product Code
- HRS
- PMA / PMN Number
- K123964
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NURSE
Narratives
THE REPORTED EVENT THAT 7 SCREWS BACKED OUT FROM THE PLATE COULD BE CONFIRMED. BASED ON INVESTIGATION, THE ROOT CAUSE WAS ATTRIBUTED TO BE USER RELATED. THE FAILURE WAS CAUSED BY INCORRECT SELECTION AND PLACEMENT OF THE SCREWS IN THE PLATE. THE X-RAYS PROVIDED SHOWED THAT 7 SCREWS WERE BACKED OUT OF THE PLATE IN THE PROXIMAL END. THESE SCREWS WERE RETURNED AND INSPECTED AND DESPITE OF THE USUAL MARKS OF USAGE (SCRATCHES IN THE TORX HEAD), NO OTHER DAMAGES WERE NOTICED. THE LOCKING THREADS ARE STILL COMPLETELY INTACT. A CLINICAL EVALUATION OF THE X-RAYS PROVIDED WAS REQUESTED AND, THE FOLLOWING WAS CONCLUDED: ¿THE GIVEN CASE PRESENTS A TYPICAL IMPLANT FAILURE CAUSED BY POOR SURGICAL TECHNIQUE (TOO SHORT LOCKING SCREWS IN THE HUMERAL HEAD AND INSUFFICIENT TIGHTENING OF THREE LOCKING SCREWS THAT WERE INSERTED IN THE PROXIMAL MAIN FRAGMENT).¿ BOTH CLINICAL EVALUATION AND INSPECTION POINT TO THE FACT THAT THE LOCKING SCREWS WERE NOT PROPERLY LOCKED IN THE PLATE. PROPER LOCKING CAN ONLY BE ACHIEVED BY THE CORRECT USE OF THE TORQUE LIMITER TO TIGHTEN THE SCREW AND PROVIDE THE APPROPRIATE TORQUE TO IT. MOREOVER THE CLINICAL EXPERT POINTED OUT THAT THE USAGE OF TOO SHORT SCREWS WHICH ULTIMATELY WAS WHAT LED TO THE FAILURE OF THE DEVICE. THE USE OF TOO SHORT SCREWS MEANS THAT THE THERE WAS NOT ENOUGH GRIP ON THE BONE SO THE SCREWS WERE NOT SECURELY ATTACHED. NOTE, AS STATED IN THE IFU (V15013): ¿4 WARNINGS AND PRECAUTIONS ¿ IMPLANT SELECTION AND SIZING: THE CORRECT SELECTION OF THE FRACTURE FIXATION APPLIANCE IS EXTREMELY IMPORTANT. FAILURE TO USE THE APPROPRIATE APPLIANCE FOR THE FRACTURE CONDITION MAY ACCELERATE CLINICAL FAILURE. FAILURE TO USE THE PROPER COMPONENT TO MAINTAIN ADEQUATE BLOOD SUPPLY AND PROVIDE RIGID FIXATION MAY RESULT IN LOOSENING, BENDING, CRACKING OR FRACTURE OF THE DEVICE AND/OR BONE. THE CORRECT IMPLANT SIZE FOR A GIVEN PATIENT CAN BE DETERMINED BY EVALUATING THE PATIENT¿S HEIGHT, WEIGHT, FUNCTIONAL DEMANDS AND ANATOMY. EVERY IMPLANT MUST BE USED IN THE CORRECT ANATOMIC LOCATION, CONSISTENT WITH ACCEPTED STANDARDS OF INTERNAL FIXATION.¿ NOTE, AS STATED IN THE OPERATIVE TECHNIQUE (AXSOS-ST-14-EN REV 1 AXSOS 3 TITANIUM PROXIMAL HUMERUS OPTECH_2015-6882): ¿NOTE: CORRECT LENGTH OF THE LOCKING SCREWS IN THE HUMERAL HEAD ARE ESSENTIAL. ¿ PARTICULARLY IN OSTEOPENIC BONE, THERE IS ONLY LIMITED SUBSTANTIAL CANCELLOUS BONE STRUCTURE CLOSE TO THE SUBCHONDRAL COMPACTA PROVIDING RELIABLE STRENGTH FOR A BONE SCREW. TOO CLOSE OF A DISTANCE OF THE SCREW TIP TO THE SUBCHONDRAL COMPACTA BEARS AN INCREASED RISK OF CUT OUT IN THE CASE OF POSTOPERATIVE BONE SINTERING. FINAL TIGHTENING OF THE LOCKING SCREW IS ALWAYS PERFORMED BY HAND USING THE 2.5NM TORQUE LIMITER (702760) IN COMBINATION WITH A SCREWDRIVER BIT T15 (705015) AND THE T-HANDLE (702427). PERFORMING FINAL TIGHTENING BY HAND WILL HELP TO PREVENT OVER-TIGHTENING OF LOCKING SCREWS, AND ALSO ENSURES THAT THESE SCREWS ARE TIGHTENED TO A TORQUE OF 2.5NM. THE DEVICE WILL CLICK WHEN THE TORQUE REACHES 2.5NM. ENSURE THAT THE SCREWDRIVER TIP IS FULLY SEATED IN THE SCREW HEAD, AND DO NOT ANGULATE THE SCREWDRIVER.¿ A REVIEW OF THE DEVICE HISTORY FOR THE REPORTED LOT DID NOT INDICATE ANY ABNORMALITIES. NO CORRECTIVE ACTIONS ARE REQUIRED AT THIS TIME. A REVIEW OF THE LABELING DID NOT INDICATE ANY ABNORMALITIES. NO INDICATIONS OF MATERIAL, MANUFACTURING OR DESIGN RELATED PROBLEMS WERE FOUND DURING THE INVESTIGATION. IF ANY FURTHER INFORMATION IS PROVIDED, THE INVESTIGATION REPORT WILL BE UPDATED.
DEVICE WILL NOT BE RETURNED. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, IT WILL BE PROVIDED ON A SUPPLEMENTAL REPORT.
CONDITION AFTER MULTIPLE PROXIMAL HUMERAL PLEXUS FRACTURE, WHICH WAS PROVIDED WITH THE AXOS HUMERUS PLATE AT THE END OF (B)(6) 2016. THROUGHOUT, THE PATIENT MOAN ABOUT PAIN IN THE AREA OF THE OPERATED ARM. IN AN X-RAY INSPECTION IT WAS SHOWN THAT A TOTAL OF 7 SCREWS WERE LOOSENED, ONE OF THESE SCREWS WAS COMPLETELY EXTRAOSSARY. CLINICALLY AND PARACLINICALLY, INFECTIONS COULD BE EXCLUDED. THE PLATE WAS REFIXED IN A REVISION SURGERY
CONDITION AFTER MULTIPLE PROXIMAL HUMERAL PLEXUS FRACTURE, WHICH WAS PROVIDED WITH THE AXOS HUMERUS PLATE AT THE END OF (B)(6) 2016. THROUGHOUT, THE PATIENT MOAN ABOUT PAIN IN THE AREA OF THE OPERATED ARM. IN AN X-RAY INSPECTION IT WAS SHOWN THAT A TOTAL OF 7 SCREWS WERE LOOSENED, ONE OF THESE SCREWS WAS COMPLETELY "EXTRAOSSARY." CLINICALLY AND PARACLINICALLY, INFECTIONS COULD BE EXCLUDED. THE PLATE WAS REFIXED IN A REVISION SURGERY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 29041 | CORTEX SCREW AXSOS 3 TI Ø3.5MM / L36MM | PLATE, FIXATION, BONE | HRS | STRYKER GMBH | V18132 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |