FDA Adverse Event
Injury
Summary report: N
FUSE ENDOSCOPY SYSTEM
MDR report key: 6244035
·
Received January 11, 2017
Report
- Report Number
- 3007591333-2017-00002
- Event Type
- Injury
- Date Received
- January 11, 2017
- Date of Event
- November 28, 2016
- Report Date
- January 11, 2017
- Manufacturer
- ENDOCHOICE, INC.
- Product Code
- FDF
- PMA / PMN Number
- K141598
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS RETURNED TO THE MANUFACTURER SERVICE CENTER. THE INVESTIGATION DID NOT IDENTIFY ANY ANOMALIES WITH THE DEVICE THAT WOULD PREVENT SUCCESSFUL CLEANING AND HIGH-LEVEL DISINFECTION WHEN PERFORMED IN ACCORDANCE WITH VALIDATED INSTRUCTIONS. TESTING SHOWED THAT ANGULATION DID NOT MEET OEM SPECIFICATIONS. REPAIR INCLUDED ADJUSTMENT OF ANGULATION ANGLES TO MEET SPECIFICATION, AND REPLACEMENT OF THE BENDING SHEATH ON THE DISTAL TIP.
Description of Event or Problem · 1
IT WAS REPORTED THAT AFTER COMPLETION OF A NORMAL COLONOSCOPY WITH NO SPECIMENS COLLECTED, THE PATIENT WAS DIAGNOSED WITH COLITIS AND RECEIVED ANTIBIOTIC TREATMENT. ACCORDING TO THE REPORTER, THE SCOPE EMPLOYED TO PERFORM THE COLONOSCOPY WAS USED 153 TIMES BETWEEN (B)(6) 2016.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 27001 | FUSE ENDOSCOPY SYSTEM | COLONOSCOPE AND ACCESSORIES | FDF | ENDOCHOICE, INC. | FUSE 1C | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR | Required Intervention |