FDA Adverse Event
Malfunction
Summary report: N
MERLIN PROGRAMMER
MDR report key: 6239676
·
Received January 10, 2017
Report
- Report Number
- 2938836-2017-02335
- Event Type
- Malfunction
- Date Received
- January 10, 2017
- Date of Event
- December 23, 2015
- Report Date
- October 11, 2016
- Manufacturer
- ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
- Product Code
- KRG
- PMA / PMN Number
- P030054
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
(B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT THROUGH MERLIN.NET TRANSMISSION, AN UNEXPECTED DECREASE IN DEVICE LONGEVITY WAS OBSERVED. ON (B)(6) 2015 REMOTE TRANSMISSION, BATTERY LONGEVITY WAS 3.7-3.9 YEARS. ON ANOTHER TRANSMISSION ON (B)(6) 2016, ESTIMATED LONGEVITY WAS 2.0 YEARS. ST. JUDE MEDICAL TECHNICAL REPRESENTATIVE PERFORMED IN-HOUSE CALCULATIONS AND EXPLAINED THAT THE DEVICE OVER-ESTIMATED THE LONGEVITY IN (B)(6) 2015. IT WAS CLOSER TO 2.0-2.75 YEARS IN (B)(6) 2015. NOW, THE PROGRAMMER DISPLAYED ESTIMATE OF 1.3 YEARS REMAINING TO ERI AND IT SHOULD BE ACCURATE FROM THIS POINT ONWARD. THE PATIENT WILL CONTINUE TO BE FOLLOWED NORMALLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 21889 | MERLIN PROGRAMMER | PACEMAKER PROGRAMMER | KRG | ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) | 3650 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 74 YR |