FDA Adverse Event Malfunction Summary report: N

MERLIN PROGRAMMER

MDR report key: 6239676 · Received January 10, 2017

Report

Report Number
2938836-2017-02335
Event Type
Malfunction
Date Received
January 10, 2017
Date of Event
December 23, 2015
Report Date
October 11, 2016
Manufacturer
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
Product Code
KRG
PMA / PMN Number
P030054
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THROUGH MERLIN.NET TRANSMISSION, AN UNEXPECTED DECREASE IN DEVICE LONGEVITY WAS OBSERVED. ON (B)(6) 2015 REMOTE TRANSMISSION, BATTERY LONGEVITY WAS 3.7-3.9 YEARS. ON ANOTHER TRANSMISSION ON (B)(6) 2016, ESTIMATED LONGEVITY WAS 2.0 YEARS. ST. JUDE MEDICAL TECHNICAL REPRESENTATIVE PERFORMED IN-HOUSE CALCULATIONS AND EXPLAINED THAT THE DEVICE OVER-ESTIMATED THE LONGEVITY IN (B)(6) 2015. IT WAS CLOSER TO 2.0-2.75 YEARS IN (B)(6) 2015. NOW, THE PROGRAMMER DISPLAYED ESTIMATE OF 1.3 YEARS REMAINING TO ERI AND IT SHOULD BE ACCURATE FROM THIS POINT ONWARD. THE PATIENT WILL CONTINUE TO BE FOLLOWED NORMALLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
21889 MERLIN PROGRAMMER PACEMAKER PROGRAMMER KRG ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) 3650

Patients

Seq Age Sex Outcome Treatment
1 74 YR