FDA Adverse Event Malfunction Summary report: N

3.0MM TAPERED ROUTER

MDR report key: 6239283 · Received January 10, 2017

Report

Report Number
0001811755-2017-00036
Event Type
Malfunction
Date Received
January 10, 2017
Date of Event
December 14, 2016
Report Date
March 21, 2017
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
HBE
UDI-DI
07613327096736
PMA / PMN Number
K143399
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A FOLLOW UP REPORT WILL BE FILED ONCE THE QUALITY INVESTIGATION IS COMPLETE. ROUTER DISCARDED.

Additional Manufacturer Narrative · 1

IT WAS INITIALLY REPORTED THAT THE ROUTER WAS DISCARDED, A PARTIAL PIECE WAS RETURNED WITH THE ATTACHMENT. INVESTIGATION RESULTS CONCLUDED THAT AS THE ROUTER WAS LOADED WITHOUT THE ATTACHMENT. THIS COULD POTENTIALLY CAUSE THE FRACTURE OF THE ROUTER. A REVIEW OF THE ASSOCIATED ATTACHMENT IFU 5407-210-768 REV-A NOTED THE FOLLOWING WARNING; "NOTE: CUTTING ACCESSORY LABELS ARE MARKED WITH A COLOR STRIPE TO FACILITATE CUTTING ACCESSORY SELECTION. TO SELECT AN APPROPRIATE CUTTING ACCESSORY, MATCH THE COLOR STRIPE ON THE CUTTING ACCESSORY LABEL WITH THE COLOR STRIPE ON THE ATTACHMENT. WHILE ALIGNING THE DOTS ON THE ATTACHMENT AND MOTOR, SLIDE THE ATTACHMENT OVER THE CUTTING ACCESSORY AND ONTO THE MOTOR UNTIL IT SNAPS INTO PLACE. WARNING: OPERATING THE MOTOR WITHOUT THE ATTACHMENT SECURELY CONNECTED AND CUTTING ACCESSORY SECURELY LOCKED MAY RESULT IN PATIENT OR HEALTHCARE STAFF INJURY."

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PRODUCT DEMONSTRATION WITH THE PI DRIVE PLUS MOTOR, THE ROUTER BROKE. IT WAS ALSO REPORTED THAT THIS EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE, AND THERE WAS NO DELAY OR ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PRODUCT DEMONSTRATION WITH THE PI DRIVE PLUS MOTOR, THE ROUTER BROKE. IT WAS ALSO REPORTED THAT THIS EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE, AND THERE WAS NO DELAY OR ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
20844 3.0MM TAPERED ROUTER DRILLS, BURRS, TREPHINES & ACCESSORIES HBE STRYKER INSTRUMENTS-KALAMAZOO 16204017 07613327096736

Patients

Seq Age Sex Outcome Treatment
1 (B)(4)