FDA Adverse Event Injury Summary report: N

POLYSORB 0 VL 5X18 BTP-1 DT

MDR report key: 6236227 · Received January 10, 2017

Report

Report Number
9612501-2017-00061
Event Type
Injury
Date Received
January 10, 2017
Date of Event
December 9, 2016
Report Date
December 12, 2016
Manufacturer
COVIDIEN
Product Code
GAM
PMA / PMN Number
K963253
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4). POST MARKET VIGILANCE (PMV) LED AN EVALUATION OF ONE DEVICE. VISUAL INSPECTION OF THE RETAINER NOTED THAT FOUR WITNESS MARKS WERE PRESENT, WHILE THE CORRECT NUMBER OF UNITS FOR THIS PRODUCT IS FIVE; THIS SERVES AS AN INDICATION THAT THE INCORRECT NUMBER OF UNITS WAS PLACED IN THE PACKAGING AT THE TIME OF MANUFACTURE. THE ROOT CAUSE OF THE OBSERVED CONDITION WAS DETERMINED TO BE A RESULT OF AN ASSEMBLY ERROR. THIS ISSUE HAS BEEN BROUGHT TO THE ATTENTION OF THE APPROPRIATE MANUFACTURING PERSONNEL AND A PROCESS ENHANCEMENT HAS BEEN IMPLEMENTED TO PREVENT THIS CONDITION FROM RECURRING. SHOULD NEW INFORMATION BECOME AVAILABLE, THE FILE WILL BE RE-OPENED AND THE INVESTIGATION SUMMARY WILL BE AMENDED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

CUSTOMER STATES: AFTER SURGERY, THEY COUNTED THE NUMBER OF THE NEEDLES AND NOTED THE NUMBER DID NOT CORRESPOND. THEY LOOKED FOR THE NEEDLE, BUT WERE NOT ABLE TO FIND IT OUT. SURGEON DECIDED TO CLOSE THE ABDOMEN AS HE/SHE THOUGHT THE NEEDLE HAD NOT BEEN CONTAINED IN THE PACKET FROM THE BEGINNING. HOWEVER, POSTOPERATIVELY THE PATIENT UNDERWENT A CT SCAN TO BE SAFE. THE EVENT OCCURRED IN USE FOR PATIENT. THE SURGICAL TIME WAS EXTENDED BY MORE THAN 30 MIN. THE STATUS OF THE PATIENT: NO PROBLEM. THERE WAS NO TISSUE DAMAGE. THE INCISION SITE WAS NOT EXTENDED. NOTHING FELL INTO THE PATIENT'S CAVITY. NO BLEEDING OCCURRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
21333 POLYSORB 0 VL 5X18 BTP-1 DT LAPAROSCOPE, GENERAL & PLASTIC SURGERY GAM COVIDIEN CLT250MG D6B2566X

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention