FDA Adverse Event Malfunction Summary report: N

BIAD XLT

MDR report key: 62352 · Received January 13, 1997

Report

Report Number
1527587-1997-00001
Event Type
Malfunction
Date Received
January 13, 1997
Date of Event
January 3, 1997
Report Date
January 9, 1997
Manufacturer
TRIONIX RESEARCH LABORATORY, INC.
Product Code
JWM
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

WHILE 67 YEAR OLD MALE WAS BEING SCANNED, HEAD 2 DETECTOR DROPPED AND TOUCHED PT, PT WAS EXAMINED AT THE HOSP, NO INJURIES WERE FOUND. APPARENTLY THE RADIUS BALL SCREW BROKE AT THE JUNCTION OF THE MOTOR, ALLOWING THE DETECTOR TO DROP DOWNWARD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BIAD XLT NUCLEAR DIAGNOSTIC IMAGING CAMERA JWM TRIONIX RESEARCH LABORATORY, INC. BIAD XLT NA

Patients

Seq Age Sex Outcome Treatment
1 67 YR Other