SMARTSET MV ENDURANCE 80G
Report
- Report Number
- 1818910-2017-10435
- Event Type
- Injury
- Date Received
- January 9, 2017
- Date of Event
- January 4, 2017
- Report Date
- January 4, 2017
- Manufacturer
- DEPUY CMW 9610921
- Product Code
- LOD
- PMA / PMN Number
- K033382
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
Narratives
IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. (B)(4). THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.
NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. A WORLDWIDE COMPLAINT DATABASE SEARCH FOUND ONE OTHER REPORTED INCIDENT(S) AGAINST 3102080 SMARTSET MV ENDURANCE 80G. A PREVIOUS DHR FOUND NO RELATED DEVIATIONS OR ANOMALIES. A WORLDWIDE COMPLAINT DATABASE SEARCH FOUND NO OTHER REPORTED INCIDENT(S) AGAINST THE REMAINING PROVIDED PRODUCT/LOT COMBINATION(S) SINCE RELEASE FOR DISTRIBUTION. BASED ON PREVIOUS INVESTIGATIONS THIS COMPLICATION OF JOINT REPLACEMENT IS UNLIKELY TO HAVE BEEN THE RESULT OF A DEVICE FAILING TO MEET REQUIRED SPECIFICATIONS. THE INFORMATION RECEIVED WILL BE RETAINED FOR POTENTIAL SERIES INVESTIGATIONS IF TRIGGERED BY TREND ANALYSIS, POST MARKET SURVEILLANCE, OR OTHER EVENTS WITHIN THE QUALITY SYSTEM. CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.
PATIENT WAS REVISED TO ADDRESS TIBIAL AND FEMORAL LOOSENING. LOOSENING OCCURRED AT THE BONE/CEMENT AND CEMENT/IMPLANT INTERFACE. CEMENT MANUFACTURED BY DEPUY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 16816 | SMARTSET MV ENDURANCE 80G | CEMENT / CEMENT ACCESSORY | LOD | DEPUY CMW 9610921 | 2803358 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 78 YR | Required Intervention |