FDA Adverse Event Malfunction Summary report: N

CALIBER INFLATION DEVICE

MDR report key: 6233303 · Received January 9, 2017

Report

Report Number
2020394-2016-01239
Event Type
Malfunction
Date Received
January 9, 2017
Date of Event
December 12, 2016
Report Date
May 24, 2017
Manufacturer
PEROUSE MEDICAL
Product Code
MAV
UDI-DI
00801741090721
PMA / PMN Number
K042449
Removal / Correction Number
Z-1891-2017
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

NO HOSPITAL/MEDICAL RECORDS OR MEDICAL IMAGES HAVE BEEN MADE AVAILABLE TO THE MANUFACTURER. ONE LABELING PHOTO HAS BEEN MADE AVAILABLE. AS THE LOT NUMBER FOR THE DEVICE WAS PROVIDED, A REVIEW OF THE DEVICE HISTORY RECORDS IS CURRENTLY BEING PERFORMED. THE DEVICE HAS BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. THE INVESTIGATION OF THE REPORTED EVENT IS CURRENTLY UNDERWAY. THE INFORMATION PROVIDED BY BARD REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BARD.

Additional Manufacturer Narrative · 1

THE DEVICE HISTORY RECORDS WERE REVIEWED WITH SPECIAL ATTENTION TO THE RAW MATERIALS, SUBASSEMBLIES, MANUFACTURING PROCESS AND QUALITY CONTROL TESTING. THIS LOT MET ALL RELEASE CRITERIA. THERE WAS NOTHING FOUND TO INDICATE THERE WAS A MANUFACTURING RELATED CAUSE FOR THIS EVENT. VISUAL/MICROSCOPIC INSPECTION: THE DEVICE APPEARED CLEANED. THE TYVEK SEAL IS INTACT. A CRACK WAS NOTED IN THE CLEAR PLASTIC OF THE PACKAGING. THE CRACK WAS NOTED ALONG THE END OF THE PACKAGING NEAR THE PRESSURE GAUGE. NO OTHER DAMAGE WAS NOTED TO THE PACKAGING. MEDICAL RECORDS REVIEW: MEDICAL RECORDS WERE NOT PROVIDED; THEREFORE, A REVIEW COULD NOT BE PERFORMED. PHOTO REVIEW: TWO DIGITAL PHOTOS WERE REVIEWED. THE FIRST PHOTO SHOWS THE LABELING OF THE DEVICE CONFIRMING IT TO BE A CALIBER INFLATION DEVICE. THE SECOND PHOTO SHOWS A CRACK IN THE PACKAGING NEAR THE PRESSURE GAUGE OF THE DEVICE. BASED ON THE PROVIDED PHOTOS, THE INVESTIGATION CAN BE CONFIRMED FOR PACKAGING ISSUE. CONCLUSION: THE DEVICE WAS RETURNED. BASED ON THE RETURN CONDITION, THE INVESTIGATION IS CONFIRMED FOR BREAK OF STERILE BARRIER DUE TO A PACKAGING ISSUE. THE DEFINITIVE ROOT CAUSE FOR THE REPORTED EVENT COULD NOT BE DETERMINED BASED UPON AVAILABLE INFORMATION. LABELING REVIEW: THE CURRENT IFU (INSTRUCTIONS FOR USE) STATES: WARNING: CONTENTS SUPPLIED STERILE USING ETHYLENE OXIDE (EO). NON-PYROGENIC. DO NOT USE IF STERILE BARRIER IS OPENED OR DAMAGED. SINGLE PATIENT USE ONLY. DO NOT REUSE, REPROCESS OR RE-STERILIZE. THE INFORMATION PROVIDED BY BARD REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BARD.

Additional Manufacturer Narrative · 1

(B)(4). MANUFACTURING REVIEW: THE DEVICE HISTORY RECORDS WERE REVIEWED WITH SPECIAL ATTENTION TO THE RAW MATERIALS, SUBASSEMBLIES, MANUFACTURING PROCESS AND QUALITY CONTROL TESTING. THIS LOT MET ALL RELEASE CRITERIA. THERE WAS NOTHING FOUND TO INDICATE THERE WAS A MANUFACTURING RELATED CAUSE FOR THIS EVENT. VISUAL/MICROSCOPIC INSPECTION: THE DEVICE APPEARED CLEANED. THE TYVEK SEAL IS INTACT. A CRACK WAS NOTED IN THE CLEAR PLASTIC OF THE PACKAGING. THE CRACK WAS NOTED ALONG THE END OF THE PACKAGING NEAR THE PRESSURE GAUGE. NO OTHER DAMAGE WAS NOTED TO THE PACKAGING. MEDICAL RECORDS REVIEW: MEDICAL RECORDS WERE NOT PROVIDED; THEREFORE, A REVIEW COULD NOT BE PERFORMED. PHOTO REVIEW: TWO DIGITAL PHOTOS WERE REVIEWED. THE FIRST PHOTO SHOWS THE LABELING OF THE DEVICE CONFIRMING IT TO BE A CALIBER INFLATION DEVICE. THE SECOND PHOTO SHOWS A CRACK IN THE PACKAGING NEAR THE PRESSURE GAUGE OF THE DEVICE. BASED ON THE PROVIDED PHOTOS, THE INVESTIGATION CAN BE CONFIRMED FOR PACKAGING ISSUE. CONCLUSION: THE DEVICE WAS RETURNED. BASED ON THE RETURN CONDITION, THE INVESTIGATION IS CONFIRMED FOR BREAK OF STERILE BARRIER DUE TO A PACKAGING ISSUE. THE DEFINITIVE ROOT CAUSE FOR THE REPORTED EVENT COULD NOT BE DETERMINED BASED UPON AVAILABLE INFORMATION. LABELING REVIEW: THE CURRENT IFU (INSTRUCTIONS FOR USE) STATES: WARNING: CONTENTS SUPPLIED STERILE USING ETHYLENE OXIDE (EO). NON-PYROGENIC. DO NOT USE IF STERILE BARRIER IS OPENED OR DAMAGED. SINGLE PATIENT USE ONLY. DO NOT REUSE, REPROCESS OR RE-STERILIZE. THE INFORMATION PROVIDED BY BARD REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BARD.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN ALLEGED CRACK WAS OBSERVED ON THE PLASTIC PACKAGING OF THE INFLATION DEVICE. THERE WAS NO REPORTED PATIENT CONTACT.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN ALLEGED CRACK WAS OBSERVED ON THE PLASTIC PACKAGING OF THE INFLATION DEVICE. THERE WAS NO REPORTED PATIENT CONTACT.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN ALLEGED CRACK WAS OBSERVED ON THE PLASTIC PACKAGING OF THE INFLATION DEVICE. THERE WAS NO REPORTED PATIENT CONTACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
18729 CALIBER INFLATION DEVICE ANGIOGRAPHIC INJECTOR AND SYRINGE MAV PEROUSE MEDICAL 15085271 00801741090721

Patients

Seq Age Sex Outcome Treatment
1