FDA Adverse Event Injury Summary report: N

20 G X 1.16 IN. BD INSYTE¿ AUTOGUARD¿ SHIELDED IV CATHETER

MDR report key: 6228267 · Received January 6, 2017

Report

Report Number
1710034-2017-00001
Event Type
Injury
Date Received
January 6, 2017
Date of Event
December 13, 2016
Report Date
January 24, 2017
Manufacturer
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
Product Code
FOZ
PMA / PMN Number
K952861
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: A SAMPLE HAS BEEN RECEIVED FOR EVALUATION. A SUPPLEMENTAL REPORT WILL BE FILED UPON COMPLETION OF THE INVESTIGATION.

Additional Manufacturer Narrative · 1

THE MANUFACTURING SITE RECEIVED ONE USED (B)(4) UNIT WITH AN OPENED PACKAGE FROM THE LOT NUMBER 6061733. THE UNIT CONSISTED OF THE NEEDLE SAFETY BARREL ASSEMBLY. A VISUAL/MICROSCOPIC EXAMINATION WAS PERFORMED AND IT WAS OBSERVED THAT THE NEEDLE WAS FULLY RETRACTED INTO THE SAFETY BARREL AND THE WHITE BUTTON WAS DEPRESSED. THE NEEDLE WAS PUSHED AND REPOSITIONED TO THE OUT POSITION; OBSERVED THE NEEDLE WAS SLIGHTLY BENT APPROXIMATELY 5/8 OF AN INCH ABOVE THE NEEDLE HUB. NO NEEDLE COVER WAS RETURNED FOR EVALUATION. A FUNCTION TEST (NEEDLE RETRACTION) WAS PERFORMED. THE NEEDLE WAS PUSHED AND REPOSITIONED TO THE OUT POSITION. THE BUTTON WAS DEPRESSED AND THE NEEDLE RETRACTED FULLY WITHIN THE SAFETY BARREL. THE RETRACTION WAS SUCCESSFUL, MEETING NO RESISTANCE. A REVIEW OF THE DEVICE HISTORY RECORD REVEALED NO ABNORMALITIES DURING THE MANUFACTURE OF THE REPORTED LOT NUMBER 6061733. THERE WERE NO REJECT ACTIVITY FINDINGS THROUGHOUT THE BUILD OF THIS LOT THAT WOULD IMPACT UPON THE QUALITY OF THE PRODUCT. THE CUSTOMER'S REPORTED DEFECT WAS NOT CONFIRMED AS THE RETURNED UNIT DID NOT DISPLAY ANY ADVERSE CHARACTERISTICS THAT WOULD CONTRIBUTE TO THE DEFECTS THE CUSTOMER EXPERIENCED. THERE WAS NO DEFINITE PHYSICAL OR MECHANICAL EVIDENCE THAT CONFIRMED AND SUPPORTED MANUFACTURING PROCESS RELATED ISSUES FOR THE DEFECT OBSERVED IN THIS INCIDENT. AN ABSOLUTE ROOT CAUSE FOR THIS INCIDENT CANNOT BE DETERMINED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE EMPLOYEE STARTED THE IV WITH THE SUSPECT DEVICE. WHEN THE EMPLOYEE PUSHED THE BUTTON TO RETRACT THE NEEDLE, IT RETRACTED AND THEN CAME OUT THE SIDE, RESULTING IN A NEEDLE STICK INJURY TO THE EMPLOYEE'S HAND.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
12419 20 G X 1.16 IN. BD INSYTE¿ AUTOGUARD¿ SHIELDED IV CATHETER SHIELDED IV CATHETER FOZ BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. 6061733

Patients

Seq Age Sex Outcome Treatment
1 Other