FDA Adverse Event Other Summary report: N

XL TEK EVOKED POTENITAL HEADBOX

MDR report key: 622690 · Received July 19, 2005

Report

Report Number
9612330-2005-00001
Event Type
Other
Date Received
July 19, 2005
Date of Event
June 14, 2005
Report Date
July 15, 2005
Manufacturer
EXCEL TECH. LTD.
Product Code
GWF
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Description of Event or Problem · 1

DURING A SPINAL CORD (THORACI TUMOR CASE-2 POST TIBIAL NERVE, RIGHT MEDIAN), PT UNDERWENT 3 EXTREMITY SSEP TESTING. PT WAS PLACED IN PRONE POSITION FOR CASE WITH IOM NEEDLE ELECTRODES PLACED TO SCALP AND BILATERAL CHEST WALL. DURING CLOSING, A 1.5 CM DIAMETER SKIN LESION WAS NOTED ON L) FOREHEAD WHERE GROUND OR FPZ ELECTRODE WAS. ALSO NOTED 4 LESIONS ON CHEST. LESIONS SITES WERE GRAY WITH RED OUTLINE.

Description of Event or Problem · 1

THE ELECTROSURGERY DEVICE CURRENT MAY HAVE ESTABLISHED AN ALTERNATE RETURN PATH THAT MAY HAVE RESULTED IN HIGH CURRENT DENSITY PASSING THROUGH THE SUBDERMAL NEEDLE ELECTRODES WITH THE POSSIBLE TISSUE HEATING AND DAMAGE RESULTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 XL TEK EVOKED POTENITAL HEADBOX EVOKED POTENTIAL DIAGNOSTIC DEVICE GWF EXCEL TECH. LTD. EP16 NA

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other