FDA Adverse Event
Other
Summary report: N
XL TEK EVOKED POTENITAL HEADBOX
MDR report key: 622690
·
Received July 19, 2005
Report
- Report Number
- 9612330-2005-00001
- Event Type
- Other
- Date Received
- July 19, 2005
- Date of Event
- June 14, 2005
- Report Date
- July 15, 2005
- Manufacturer
- EXCEL TECH. LTD.
- Product Code
- GWF
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Description of Event or Problem · 1
DURING A SPINAL CORD (THORACI TUMOR CASE-2 POST TIBIAL NERVE, RIGHT MEDIAN), PT UNDERWENT 3 EXTREMITY SSEP TESTING. PT WAS PLACED IN PRONE POSITION FOR CASE WITH IOM NEEDLE ELECTRODES PLACED TO SCALP AND BILATERAL CHEST WALL. DURING CLOSING, A 1.5 CM DIAMETER SKIN LESION WAS NOTED ON L) FOREHEAD WHERE GROUND OR FPZ ELECTRODE WAS. ALSO NOTED 4 LESIONS ON CHEST. LESIONS SITES WERE GRAY WITH RED OUTLINE.
Description of Event or Problem · 1
THE ELECTROSURGERY DEVICE CURRENT MAY HAVE ESTABLISHED AN ALTERNATE RETURN PATH THAT MAY HAVE RESULTED IN HIGH CURRENT DENSITY PASSING THROUGH THE SUBDERMAL NEEDLE ELECTRODES WITH THE POSSIBLE TISSUE HEATING AND DAMAGE RESULTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | XL TEK EVOKED POTENITAL HEADBOX | EVOKED POTENTIAL DIAGNOSTIC DEVICE | GWF | EXCEL TECH. LTD. | EP16 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |