FDA Adverse Event Malfunction Summary report: N

AMIS TERMINAL CUP IMPACTOR FOR METAL BACK CC

MDR report key: 6224856 · Received January 5, 2017

Report

Report Number
3005180920-2016-00711
Event Type
Malfunction
Date Received
January 5, 2017
Date of Event
December 7, 2016
Report Date
February 3, 2017
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LXH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ON 18 JANUARY 2017 THE (B)(6) PERFORMED A VISUAL INSPECTION OF THE RETRIEVED ITEM AND COMMENTED AS FOLLOWS: ACCORDING TO OUR VISUAL INSPECTION THERE IS IMPINGEMENT BETWEEN : TEETH OF THE CUP AND THE TERMINAL RIM. INTERNAL SURFACE OF THE CUP AND BOTTOM OF THE CUP IMPACTOR. FOR THIS REASON, THESE INSTRUMENT ARE NOT COMPATIBLE WITH VERSAFIT CUP TG. 42. THIS CUP CAN BE IMPLANTED USING PRECIMED/GREATBATCH/HPF AMIS CUP IMPACTOR.

Additional Manufacturer Narrative · 1

BATCH REVIEWS PERFORMED ON 29 DECEMBER 2016. LOT. 1311795: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 20 SEPTEMBER 2013. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, NO SIMILAR EVENTS HAVE BEEN REPORTED ON ITEMS OF THE SAME LOT. VERSAFITCUP CC TRIO ACETABULAR SHELL Ø 42, CODE 01.26.45.0042, LOT. 146525 (K122911) (B)(4) ITEMS MANUFACTURED AND RELEASED ON 21 NOVEMBER 2014. EXPIRATION DATE: 2019-10-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. NOT YET ANALYZED.

Description of Event or Problem · 1

THE TERMINAL CUP IMPACTOR AND OFFSET IMPACTOR JAMMED ONTO THE DEFINITIVE CUP, AND IT WAS THEN UNABLE TO BE RELEASED TO STAY IN SITU. IT HAD TO BE REMOVED, DURING ATTEMPTS BY THE SURGEON TO SEPARATE THE CUP FROM THE INSTRUMENTS, THESE FELL TO THE FLOOR. IT APPEARED THAT THESE INSTRUMENTS ARE NOT COMPATIBLE IN THIS SMALL SIZE, AS MANY SCRATCHES TO THE INTERNAL RIM OF THE CUP OCCURED. AS A RESULT, THIS CUP COULDN'T HAVE BEEN USED ON THE PATIENT, THE SURGEON WENT UP TO A SIZE 44MM CUP WITH HOLES AND USED THE STRAIGHT IMPACTOR WITHOUT ANY ISSUES. NO CRITICAL DELAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
10097 AMIS TERMINAL CUP IMPACTOR FOR METAL BACK CC SURGICAL INSTRUMENT FOR HIP LXH MEDACTA INTERNATIONAL SA 1311795

Patients

Seq Age Sex Outcome Treatment
1 50 YR Other