AMIS TERMINAL CUP IMPACTOR FOR METAL BACK CC
Report
- Report Number
- 3005180920-2016-00711
- Event Type
- Malfunction
- Date Received
- January 5, 2017
- Date of Event
- December 7, 2016
- Report Date
- February 3, 2017
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- LXH
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
ON 18 JANUARY 2017 THE (B)(6) PERFORMED A VISUAL INSPECTION OF THE RETRIEVED ITEM AND COMMENTED AS FOLLOWS: ACCORDING TO OUR VISUAL INSPECTION THERE IS IMPINGEMENT BETWEEN : TEETH OF THE CUP AND THE TERMINAL RIM. INTERNAL SURFACE OF THE CUP AND BOTTOM OF THE CUP IMPACTOR. FOR THIS REASON, THESE INSTRUMENT ARE NOT COMPATIBLE WITH VERSAFIT CUP TG. 42. THIS CUP CAN BE IMPLANTED USING PRECIMED/GREATBATCH/HPF AMIS CUP IMPACTOR.
BATCH REVIEWS PERFORMED ON 29 DECEMBER 2016. LOT. 1311795: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 20 SEPTEMBER 2013. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, NO SIMILAR EVENTS HAVE BEEN REPORTED ON ITEMS OF THE SAME LOT. VERSAFITCUP CC TRIO ACETABULAR SHELL Ø 42, CODE 01.26.45.0042, LOT. 146525 (K122911) (B)(4) ITEMS MANUFACTURED AND RELEASED ON 21 NOVEMBER 2014. EXPIRATION DATE: 2019-10-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. NOT YET ANALYZED.
THE TERMINAL CUP IMPACTOR AND OFFSET IMPACTOR JAMMED ONTO THE DEFINITIVE CUP, AND IT WAS THEN UNABLE TO BE RELEASED TO STAY IN SITU. IT HAD TO BE REMOVED, DURING ATTEMPTS BY THE SURGEON TO SEPARATE THE CUP FROM THE INSTRUMENTS, THESE FELL TO THE FLOOR. IT APPEARED THAT THESE INSTRUMENTS ARE NOT COMPATIBLE IN THIS SMALL SIZE, AS MANY SCRATCHES TO THE INTERNAL RIM OF THE CUP OCCURED. AS A RESULT, THIS CUP COULDN'T HAVE BEEN USED ON THE PATIENT, THE SURGEON WENT UP TO A SIZE 44MM CUP WITH HOLES AND USED THE STRAIGHT IMPACTOR WITHOUT ANY ISSUES. NO CRITICAL DELAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 10097 | AMIS TERMINAL CUP IMPACTOR FOR METAL BACK CC | SURGICAL INSTRUMENT FOR HIP | LXH | MEDACTA INTERNATIONAL SA | 1311795 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR | Other |