FDA Adverse Event Malfunction Summary report: N

COOK CELECT JUGULAR VENA CAVA FILTER SET

MDR report key: 6224392 · Received January 5, 2017

Report

Report Number
3002808486-2016-01613
Event Type
Malfunction
Date Received
January 5, 2017
Date of Event
January 15, 2010
Report Date
May 25, 2017
Manufacturer
WILLIAM COOK EUROPE
Product Code
DTK
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
ATTORNEY

Narratives

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER E2016032. WILLIAM COOK EUROPE APS (MANUFACTURER) IS SUBMITTING THIS REPORT ON BEHALF OF COOK MEDICAL INCORPORATED (CMI) (IMPORTER). (B)(4). G1) NAME AND ADDRESS FOR IMPORTER SITE: COOK MEDICAL INCORPORATED (CMI) 400 DANIELS WAY BLOOMINGTON, IN 47404 REGISTRATION NO.: 3005580113. G5) 510(K) K073374. THE EVENT IS CURRENTLY UNDER INVESTIGATION. A SUPPLEMENTAL REPORT WILL BE PROVIDED UPON CONCLUSION. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803. THIS REPORT IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT MADE IN IT IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED; THAT A DEATH OR SERIOUS INJURY OCCURRED; OR THAT ANY COOK DEVICE CAUSED OR CONTRIBUTED TO; OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER E2016032. WILLIAM COOK EUROPE APS (MANUFACTURER) IS SUBMITTING THIS REPORT ON BEHALF OF COOK MEDICAL INCORPORATED (CMI) (IMPORTER) H11) CORRECTED DATA BASED ON NEW INFORMATION RECEIVED: B1: ADVERSE EVENT TO PRODUCT PROBLEM. H1: SERIOUS INJURY TO MALFUNCTION. IT HAS NOT BEEN POSSIBLE TO FURTHER INVESTIGATE OR EVALUATE THIS ALLEGED EVENT BASED ON THE LIMITED INFORMATION PROVIDED TO DATE VIA THE OPERATIVE NOTE STATING "MIGRATION, TILT, VENA CAVA PERFORATION, EMBEDDED". COOK WILL REOPEN ITS INVESTIGATION IF FURTHER INFORMATION IS RECEIVED WARRANTING SUPPLEMENTATION IN ACCORDANCE WITH 21 C.F.R. 803.56. FILTER MIGRATION IS A KNOWN POTENTIAL COMPLICATION OF VENA CAVA FILTERS. CRANIAL (INCLUDING TO THE HEART AND LUNGS) AND CAUDAL MIGRATIONS HAVE BEEN REPORTED. AMONG OTHER CAUSES, MIGRATION MAY BE ASSOCIATED WITH FILTER PLACEMENT IN IVCS WITH DIAMETERS OTHER THAN THOSE SPECIFIED IN THE INSTRUCTIONS FOR USE, IMPROPER DEPLOYMENT, DEPLOYMENT INTO CLOTS, DISLODGEMENT DUE TO LARGE CLOT BURDENS, AND (OR) PROCEDURES THAT INVOLVE OTHER DEVICES BEING PASSED THROUGH AN IN SITU FILTER. MANIPULATION IN THE AREA OF THE FILTER IMPLANT MAY CAUSE MIGRATION OR CONTRIBUTE TO CHANGES IN THE FILTER CONFIGURATION AND PLACEMENT. VENA CAVA WALL PERFORATION IS A KNOWN POTENTIAL COMPLICATION OF VENA CAVA FILTERS. BOTH SYMPTOMATIC AND ASYMPTOMATIC EVENTS HAVE BEEN REPORTED. AMONG OTHER CAUSES, VENA CAVA WALL PERFORATION MAY INADVERTENTLY BE INITIATED BY IMPROPER DEPLOYMENT, EXCESSIVE FORCE OR MANIPULATIONS NEAR AN IMPLANTED FILTER (E.G., A SURGICAL PROCEDURE IN THE VICINITY OF A FILTER) AND (OR) PROCEDURES THAT INVOLVE OTHER DEVICES BEING PASSED THROUGH AN IN SITU FILTER. FILTER PERFORATION OF THE VENA CAVA WALL IS A KNOWN RISK REPORTED IN THE PUBLISHED SCIENTIFIC LITERATURE. ALSO, PUBLISHED SCIENTIFIC LITERATURE DESCRIBES THAT MANIPULATION IN THE AREA OF FILTER PLACEMENT COULD CONTRIBUTE TO CHANGES TO THE FILTER CONFIGURATION AND PLACEMENT THEREBY POTENTIALLY INITIATE PERFORATION OF THE VENA CAVA WALL. THERE IS A CURRENT DEBATE IN THE PUBLISHED SCIENTIFIC LITERATURE ON A DIFFERENTIATION BETWEEN IVC WALL PERFORATION WITH AND WITHOUT CLINICAL SEQUELAE. E.G. FILTER LEGS MAY BE OUTSIDE THE CONTRAST LUMEN ON IMAGING WITHOUT ACTUALLY PERFORATING THE IVC WALL (KNOWN AS TENTING) AND WITH NO CLINICAL SEQUELAE. IN CONTRAST, PERFORATION OF ADJACENT ORGANS IS REPORTED WITH CLINICAL SEQUELAE. FILTER TILT IS A KNOWN POTENTIAL COMPLICATION OF VENA CAVA FILTERS. AMONG OTHER CAUSES, FILTER TILT MAY BE ASSOCIATED WITH PLACEMENT IN IVCS WITH DIAMETERS LARGER THAN THOSE SPECIFIED IN THESE INSTRUCTIONS FOR USE; IMPROPER DEPLOYMENT; MANIPULATIONS NEAR AN IMPLANTED FILTER (E.G., A SURGICAL PROCEDURE IN THE VICINITY OF A FILTER) AND (OR) PROCEDURES THAT INVOLVE DEVICES PASSING THROUGH A FILTER, OR A FAILED RETRIEVAL ATTEMPT. EXCESSIVE FILTER TILT MAY RESULT IN LOSS OF FILTER EFFICIENCY. FILTER TILT IS A KNOWN RISK IN RELATION TO FILTER IMPLANT REPORTED IN THE PUBLISHED SCIENTIFIC LITERATURE. FILTER TILT MAY HAPPEN DURING PLACEMENT OR DURING IMPLANTING PERIOD. NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS AND NOTHING INDICATES THAT THE FILTER DID NOT PERFORM AS INTENDED, E.G. INTENDED FOR THE PREVENTION OF RECURRENT PULMONARY EMBOLISM (PE) VIA PLACEMENT IN THE VENA CAVA.

Description of Event or Problem · 0

THIS ADDITIONAL INFORMATION RECEIVED ON 02/07/2017 AS FOLLOWS: PLAINTIFF ALLEGEDLY RECEIVED AN IMPLANT ON (B)(6) 2009 VIA THE JUGULAR VEIN AS A PROPHYLAXIS FOR DVT. THE PLAINTIFF THEN HAD A SUCCESSFUL REMOVAL PERFORMED ON (B)(6) 2010 VIA RIGHT INTERNAL JUGULAR ACCESS. OPERATING PHYSICIAN NOTED THAT "WE ATTEMPTED TO SNARE THE HOOK OF THE IVC FILTER, HOWEVER, AS EXPECTED IT WAS EMBEDDED IN THE WALL." THEREFORE, A DIFFERENT DIAMETER (16 FRENCH) SHEATH WAS USED AND THE "FILTER EASILY COLLAPSED INTO THE SHEATH WITHOUT ANY SIGNIFICANT RETRACTION." PLAINTIFF IS ALLEGING MIGRATION, TILT, VENA CAVA PERFORATION, OTHER: EMBEDDED.

Additional Manufacturer Narrative · 1

(B)(4). CATALOG#: UNKNOWN BUT REFERRED TO AS A COOK CELECT FILTER. SINCE CATALOG# IS UNKNOWN THE 510(K) COULD BE EITHER K061815, K073374 OR K090140. (B)(4). INVESTIGATION IS STILL IN PROGRESS. (B)(4). CATALOG#: UNKNOWN BUT REFERRED TO AS A COOK CELECT FILTER. SINCE CATALOG# IS UNKNOWN THE 510(K) COULD BE EITHER K061815, K073374 OR K090140. (B)(4). INVESTIGATION IS STILL IN PROGRESS.

Additional Manufacturer Narrative · 1

(B)(4). CATALOG#: UNKNOWN BUT REFERRED TO AS A COOK CELECT FILTER. SINCE CATALOG# IS UNKNOWN THE 510(K) COULD BE EITHER K061815, K073374 OR K090140. (B)(4). IT HAS NOT BEEN POSSIBLE TO INVESTIGATE OR EVALUATE THIS ALLEGED EVENT BASED ON THE LIMITED INFORMATION PROVIDED TO DATE. COOK WILL REOPEN ITS INVESTIGATION IF FURTHER INFORMATION IS RECEIVING WARRANTING SUPPLEMENTATION IN ACCORDANCE WITH 21 C.F.R. 803.56.

Description of Event or Problem · 1

DESCRIPTION ACCORDING TO SHORT FORM COMPLAINT FILED: IT IS ALLEGED THAT "[PT] RECEIVED A COOK CELECT FILTER ON (B)(6) 2009". PATIENT OUTCOME: IT IS ALLEGED THAT [PT] WAS INJURED WITHOUT FURTHER EXPLANATION. HOSPITAL AND MEDICAL RECORDS HAVE BEEN REQUESTED BUT NOT YET PROVIDED.

Description of Event or Problem · 1

DESCRIPTION ACCORDING TO SHORT FORM COMPLAINT FILED: IT IS ALLEGED THAT "[PT] RECEIVED A COOK CELECT FILTER ON (B)(6) 2009". PATIENT OUTCOME: IT IS ALLEGED THAT [PT] WAS INJURED WITHOUT FURTHER EXPLANATION. HOSPITAL AND MEDICAL RECORDS HAVE BEEN REQUESTED BUT NOT YET PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
7743 COOK CELECT JUGULAR VENA CAVA FILTER SET DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR DTK WILLIAM COOK EUROPE

Patients

Seq Age Sex Outcome Treatment
1 65 YR Life Threatening