FDA Adverse Event Malfunction Summary report: N

INTERSTIM II

MDR report key: 6224107 · Received January 4, 2017

Report

Report Number
3004209178-2017-00249
Event Type
Malfunction
Date Received
January 4, 2017
Report Date
February 16, 2017
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
EZW
PMA / PMN Number
P970004
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

A HEALTHCARE PROFESSIONAL VIA A MANUFACTURER REPRESENTATIVE REPORTED SEEING A CONSUMER FOR REVIEW AND THE DEVICE HAD BATTERY LIFE OF 4-5 MONTHS WITH A CAPACITY OF 86%-100%, WHICH DID NOT SEEM CONSISTENT. THE CONSUMER WAS IMPLANTED IN (B)(6) 2015. IT WAS NOTED THAT CYCLING WAS NOT USED. PROGRAMMING WAS C+2- AT 1.7 V SINCE (B)(6) 2016. THE CONSUMER WAS SEEN ON (B)(6) 2016 AND PROGRAMMING CHANGED TO 0+3- AS EFFICACY HAD DECREASED; THE CONSUMER FELT IN SADDLE. NO PROBLEM WITH THE LEAD IMPEDANCE AT THIS VISIT; HOWEVER, IT WAS NOTED THAT BACK IN (B)(6) 0+1- <(> <<)>50, BUT NOTHING ELSE.

Description of Event or Problem · 1

ADDITIONAL INFORMATION RECEIVED FROM THE REPRESENTATIVE REPORTED THAT THE NURSE INDICATED THAT TREATMENT WAS WORKING WELL. LEAD IMPEDANCE WAS CHECKED AMP 2 PW 300, AND BATTERY LIFE SHOWED 63-74 MONTHS OF SERVICE REMAINING.

Description of Event or Problem · 1

ADDITIONAL INFORMATION RECEIVED FROM THE REPRESENTATIVE REPORTED THE CONSUMER¿S RELEVANT MEDICAL HISTORY TO INCLUDE PREVIOUS SPINAL SURGERY. THE REPRESENTATIVE WOULD BE BRINGING THE CONSUMER BACK UP TO CLINIC TO CARRY OUT THE CHECKS, AND NOTED TO HAVE LISTED THE CONSUMER FOR A CHANGE OF IMPLANT IN THE MEANTIME, AS THE WAITING LIST IS FOUR (4) MONTHS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
5308 INTERSTIM II STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE EZW MEDTRONIC PUERTO RICO OPERATIONS CO. 3058

Patients

Seq Age Sex Outcome Treatment
1