FDA Adverse Event
Malfunction
Summary report: N
EV4542
MDR report key: 621607
·
Received July 7, 2005
Report
- Report Number
- 621607
- Event Type
- Malfunction
- Date Received
- July 7, 2005
- Date of Event
- May 23, 2005
- Report Date
- July 7, 2005
- Manufacturer
- PACE MEDICAL INC
- Product Code
- DRO
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- PA, US
Narratives
Description of Event or Problem · 1
DURING CODE SITUATION, NURSE TURNED OFF PACEMAKER TO DETERMINE BASIC RATE. WHEN NURSE ATTEMPTED TO TURN PACER BACK ON, PACER WOULD NOT FUNCTION. FOLLOW UP REVEALS THIS PROBLEM WAS ABLE TO BE DUPLICATED. THE DEVICE DID NOT TOLERATE BEING MOMENTARILY SWITCHED OFF AND ON. IF IT WAS ALLOWED TO SIT FOR A WHILE IN THE OFF MODE, IT WOULD RETURN TO NORMAL OPERATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | EV4542 | DEMAND PACEMAKER | DRO | PACE MEDICAL INC | EV4542 | NONE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |