COMPREHENSIVE REVERSE SHOULDER-MINI BASEPLATE
Report
- Report Number
- 0001825034-2016-05210
- Event Type
- Injury
- Date Received
- December 30, 2016
- Date of Event
- November 28, 2016
- Report Date
- May 31, 2018
- Manufacturer
- BIOMET ORTHOPEDICS
- Product Code
- KWS
- PMA / PMN Number
- PK120121
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
Narratives
REPORTED EVENT WAS CONFIRMED BY REVIEW OF THE PROVIDED OP NOTES AND X-RAYS. PER THE REVISION OPERATIVE NOTES, THE PATIENT HAD MECHANICAL FAILURE WITH SEVERE BONE LOSS. THE GLENOSPHERE AND BASEPLATE WERE LOOSE WITH A LARGE DEFECT. PER THE CT SCAN RESULTS, THERE WAS SUPERIOR DISPLACEMENT OF THE SCAPULAR COMPONENT WITH ASSOCIATED 18MM BONY FRAGMENT SUGGESTIVE OF A MODERATELY DISPLACED FRACTURE FROM THE ANTERIOR ASPECT OF THE GLENOID AND SCAPULAR SPINE. THERE APPEARS TO BE A NONDISPLACED FRACTURE THROUGH THE BASE OF THE CORACOID PROCESS AS WELL. PER THE X-RAY REVIEW, "COMPLAINT SUGGESTS LOOSENING OF ALL THE SCREWS. THIS COULD BE IDENTIFIED RADIOGRAPHICALLY AS PERIPROSTHETIC LUCENCIES SURROUNDING THE SCREWS. THE ONLY IMAGES THAT SHOW THE SCREWS ARE 1, 2, AND 5. NONE OF THESE IMAGES REVEAL THIS TO BE CONFIRMED. DHR WAS REVIEWED AND NO DISCREPANCIES WERE FOUND. ROOT CAUSE WAS UNABLE TO BE DETERMINED AS THE NECESSARY INFORMATION TO ADEQUATELY INVESTIGATE THE REPORTED EVENT WAS NOT PROVIDED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. (B)(4). CONCOMITANT MEDICAL PRODUCTS - BIOMET COMPREHENSIVE HUMERAL TRAY WITH LOCKING RING, CATALOG#: 115370, LOT #: 910020. BIOMET COMPREHENSIVE MINI HUMERAL STEM, CATALOG: #113630, LOT#: 836390. BIOMET HUMERAL BEARING, CATALOG #: XL-115363, LOT#502530. THIS REPORT IS NUMBER 4 OF 5 MDRS FILED FOR THE SAME PATIENT (REFERENCE 1825034-2016-03745 / 03746 / 03747, 05210, 05226).
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
PATIENT UNDERWENT A RIGHT SHOULDER REVISION PROCEDURE APPROXIMATELY FOUR MONTHS POST-IMPLANTATION DUE TO ALLEGED PAIN, SWELLING, LOOSE SCREWS, AND BONE FRACTURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 865430 | COMPREHENSIVE REVERSE SHOULDER-MINI BASEPLATE | PROSTHESIS, SHOULDER | KWS | BIOMET ORTHOPEDICS | NI | 374170 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR | Hospitalization| R |