FDA Adverse Event Malfunction Summary report: N

DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 6213905 · Received December 29, 2016

Report

Report Number
3004753838-2016-63613
Event Type
Malfunction
Date Received
December 29, 2016
Date of Event
December 7, 2016
Report Date
December 7, 2016
Manufacturer
DEXCOM, INC.
Product Code
MDS
UDI-DI
00386270000262
PMA / PMN Number
P120005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4) DESCRIBE EVENT OR PROBLEM - ADDITIONAL, DEVICE AVAILABLE FOR EVALUATION - ADDITIONAL, ADDITIONAL INFORMATION/DEVICE EVALUATION, DEVICE EVALUATED BY MANUFACTURER - ADDITIONAL, EVENT PROBLEM AND EVALUATION CODES - ADDITIONAL.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

THE COMPLAINT RECEIVER WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. THE RECEIVER LOG WAS REVIEWED FIRMWARE ERRORS WERE OBSERVED. THE REPORTED EVENT OF AN ERROR ICON DISPLAY WAS CONFIRMED DURING LOG REVIEW. A ROOT CAUSE COULD NOT BE DETERMINED.

Description of Event or Problem · 1

DEXCOM WAS MADE AWARE ON (B)(6) 2016, THAT ON (B)(6) 2016, THE RECEIVER DISPLAYED AN ERROR 121. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WERE RETURNED FOR EVALUATION. THE COMPLAINT COULD NOT BE CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
861933 DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM CONTINUOUS GLUCOSE MONITOR MDS DEXCOM, INC. MT22719-PNK 5214663 00386270000262

Patients

Seq Age Sex Outcome Treatment
1 13 YR