METRIX AIR
Report
- Report Number
- 1000113657-2016-02118
- Event Type
- Malfunction
- Date Received
- December 29, 2016
- Date of Event
- December 8, 2016
- Report Date
- December 29, 2016
- Manufacturer
- TRIVIDIA HEALTH, INC.
- Product Code
- NBW
- UDI-DI
- 00021292007447
- PMA / PMN Number
- K150052
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT
- Reporter Occupation
- OTHER
Narratives
(B)(4). RETURNED METER AND RETURNED TEST STRIPS EVALUATED WITH NO DEFECTS FOUND. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: USER HAD AN INACCURATE REFERENCE. (B)(4).
CONSUMER REPORTED COMPLAINT FOR LOW BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM RESULTS OBTAINED OF 60, 50 AND 232 MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 85 - 119 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2016, A BACK TO BACK BLOOD TEST WAS PERFORMED BY THE CUSTOMER FASTING AND PRODUCED TEST RESULTS OF 60 MG/DL AND E-2 USING TRUEMETRIX AIR METER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 09/28/2017 AND OPEN VIAL DATE WAS UNDISCLOSED. THE CUSTOMER STATED HE IS ON INSULIN. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: (B)(6). MEMORY CONCERNS: 50, 232, 153 AND 330.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 863915 | METRIX AIR | BLOOD GLUCOSE SYSTEM | NBW | TRIVIDIA HEALTH, INC. | METRIX AIR | MT1693 | 00021292007447 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR | SECOND THERAPY |