FDA Adverse Event Malfunction Summary report: N

TRUEMETRIX

MDR report key: 6208996 · Received December 28, 2016

Report

Report Number
1000113657-2016-02099
Event Type
Malfunction
Date Received
December 28, 2016
Date of Event
December 3, 2016
Report Date
January 17, 2017
Manufacturer
TRIVIDIA HEALTH, INC.
Product Code
NBW
UDI-DI
00050428375105
PMA / PMN Number
K140100
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). RETURNED METER AND RETURNED TEST STRIPS EVALUATED WITH NO DEFECTS FOUND. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: USER'S TEST STRIP HAD POOR STORAGE (KITCHEN). TEST STRIP UDI# (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). PRODUCT NOT YET RETURNED FOR EVALUATION. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: USER'S TEST STRIP HAD POOR STORAGE. TEST STRIP UDI# (B)(4).

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. WIFE IS CALLING ON BEHALF OF THE CUSTOMER. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM RESULTS OBTAINED OF 174, 171 AND 116 MG/DL. 174 AND 171 ARE NON-FASTING RESULTS. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 60 - 90 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2016, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE PRODUCT IS NOT STORED ACCORDING TO SPECIFICATION AND IS STORED IN THE KITCHEN. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 09/14/2017 AND OPEN VIAL DATE AT TIME OF CALL IS A MONTH AND A HALF. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: A 174MG/DL, (B)(6) 2016 02:01 PM, FASTING: NO. A 171MG/DL, (B)(6) 2016 02:00 PM, FASTING: NO. A 116MG/DL, (B)(6) 2016 12:16 PM, FASTING: YES. A 188MG/DL, (B)(6) 2016 01:10 AM, FASTING: YES. A 187MG/DL, (B)(6) 2016 01:10 AM, FASTING: YES. MEMORY CONCERNS: 174, 171, 188, 187MG/DL.

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. WIFE IS CALLING ON BEHALF OF THE CUSTOMER. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM RESULTS OBTAINED OF 174, 171 AND 116 MG/DL. 174 AND 171 ARE NON-FASTING RESULTS. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 60 - 90 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2016, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE PRODUCT IS NOT STORED ACCORDING TO SPECIFICATION AND IS STORED IN THE KITCHEN. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 09/14/2017 AND OPEN VIAL DATE AT TIME OF CALL IS A MONTH AND A HALF. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: 174MG/DL (B)(6) 2016 02:01 PM FASTING: NO; 171MG/DL (B)(6) 2016 02:00 PM FASTING: NO; 116MG/DL (B)(6) 2016 12:16 PM FASTING: YES; 188MG/DL (B)(6) 2016 01:10 AM FASTING: YES; 187MG/DL (B)(6) 2016 01:10 AM FASTING: YES. MEMORY CONCERNS: 174 171 188 187MG/DL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
858241 TRUEMETRIX BLOOD GLUCOSE SYSTEM NBW TRIVIDIA HEALTH, INC. TRUEMETRIX MT1666 00050428375105

Patients

Seq Age Sex Outcome Treatment
1 0 YR SECOND THERAPY