FDA Adverse Event Injury Summary report: N

PACKAGE, VISUM LED 2 SURGICAL LIGHT

MDR report key: 6204711 · Received December 27, 2016

Report

Report Number
0008010153-2016-00108
Event Type
Injury
Date Received
December 27, 2016
Date of Event
December 5, 2016
Report Date
May 12, 2017
Manufacturer
STRYKER-COMMUNICATIONS
Product Code
FSY
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

IT WAS REPORTED THAT THE HANDLE COVER DISENGAGED FROM THE VISUM LED 2 LIGHT. THE COVER WAS RETURNED TO STRYKER FLOWER MOUND. AFTER INVESTIGATION, IT WAS DETERMINED THAT THE O-RING, A COMPONENT OF THE VISUM LED LIGHT HANDLE ASSEMBLY, WAS MISSING AND THAT THE COVER ITSELF FUNCTIONED AS INTENDED. IT WAS ALSO CONFIRMED THAT THE HANDLE WAS VERIFIED TO BE MANUFACTURED CORRECTLY. THE VISUM LED LIGHT HANDLE O-RING FORCES THE COVER INTO POSITION SO IT CORRECTLY LATCHES. THE MISSING O-RING IS THE ROOT CAUSE OF THE FAILURE. ADDITIONALLY, WHEN THE INITIAL EMDR WAS FILED, IT WAS FILED UNDER THE REGISTRATION NUMBER OF (B)(4) AND IS CORRECTLY FILED IN THE SUPPLEMENTAL WITH THE REGISTRATION NUMBER OF (B)(4).

Additional Manufacturer Narrative · 1

IT WAS REPORTED THAT THE HANDLE COVER ALLEGEDLY FELL OFF ON TO THE SURGICAL FIELD AND THE PROCEDURE WAS DISCONTINUED. THERE WAS A REPORTED CUT FOR INSERTION OF A CATHETER TO THE PATIENT AND THE INCIDENT IS BEING INVESTIGATED. ONCE THE INVESTIGATION IS COMPLETE, A SUPPLEMENTAL WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE HANDLE COVER ALLEGEDLY FELL OFF ON TO THE SURGICAL FIELD AND THE PROCEDURE WAS DISCONTINUED. THERE WAS A REPORTED CUT FOR INSERTION OF A CATHETER TO THE PATIENT AND THE INCIDENT IS BEING INVESTIGATED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE HANDLE COVER ALLEGEDLY FELL OFF ON TO THE SURGICAL FIELD AND THE PROCEDURE WAS DISCONTINUED. THERE WAS A REPORTED CUT FOR INSERTION OF A CATHETER TO THE PATIENT AND THE INCIDENT IS BEING INVESTIGATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
855280 PACKAGE, VISUM LED 2 SURGICAL LIGHT SURGICAL LIGHT FSY STRYKER-COMMUNICATIONS

Patients

Seq Age Sex Outcome Treatment
1 Other