PLATINIUM
Report
- Report Number
- 1000165971-2016-00851
- Event Type
- Malfunction
- Date Received
- December 23, 2016
- Date of Event
- November 23, 2016
- Report Date
- June 26, 2018
- Manufacturer
- SORIN GROUP ITALIA S.R.L. - CRM FACILITY
- Product Code
- DTB
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHYSICIAN
Narratives
REPORTEDLY, THE SECOND HARDWARE RESET TO CORRECT THE ESTIMATION OF THE RESIDUAL LONGEVITY DISPLAYED BY THE PROGRAMMER WILL BE PERFORMED ON 1 JUNE 2018.
THE HARDWARE RESET PROCEDURE WAS SUCCESSFULLY PERFORMED ON (B)(6) 2018.
REPORTEDLY, AN INTEGRITY SYSTEM ALERT (WARNING [A3]) WAS RAISED ON (B)(6) 2016 UPON INTERROGATION OF THE SUBJECT ICD, IMPLANTED ON (B)(6) 2016 WITH A REPLACEMENT OF THE RIGHT VENTRICULAR LEAD. SINCE THEN, THE ICD WAS RE-INTERROGATED BUT THE ALERT DISAPPEARED.
REPORTEDLY, AN INTEGRITY SYSTEM ALERT (WARNING ) WAS RAISED ON (B)(6) 2016 UPON INTERROGATION OF THE SUBJECT ICD, IMPLANTED ON (B)(6) 2016 WITH A REPLACEMENT OF THE RIGHT VENTRICULAR LEAD. SINCE THEN, THE ICD WAS RE-INTERROGATED BUT THE ALERT DISAPPEARED.
REPORTEDLY, AN INTEGRITY SYSTEM ALERT (WARNING [A3]) WAS RAISED ON (B)(6) 2016 UPON INTERROGATION OF THE SUBJECT ICD, IMPLANTED ON (B)(6) 2016 WITH A REPLACEMENT OF THE RIGHT VENTRICULAR LEAD. SINCE THEN, THE ICD WAS RE-INTERROGATED BUT THE ALERT DISAPPEARED.
REPORTEDLY, AN INTEGRITY SYSTEM ALERT (WARNING) WAS RAISED ON (B)(6) 2016 UPON INTERROGATION OF THE SUBJECT ICD, IMPLANTED ON (B)(6) 2016 WITH A REPLACEMENT OF THE RIGHT VENTRICULAR LEAD. SINCE THEN, THE ICD WAS RE-INTERROGATED BUT THE ALERT DISAPPEARED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 852663 | PLATINIUM | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, CARDIAC RESYNCHRONIZATION | DTB | SORIN GROUP ITALIA S.R.L. - CRM FACILITY | PLATINIUM SONR CRT-D 1841 | S0122 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |