FDA Adverse Event
Malfunction
Summary report: N
DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM
MDR report key: 6201615
·
Received December 22, 2016
Doing adverse event research by hand?
The Research Assistant searches MAUDE for you, groups reports by failure mode, and cites every record.
Free to try · every answer cites its records
Report
- Report Number
- 3004753838-2016-81714
- Event Type
- Malfunction
- Date Received
- December 22, 2016
- Date of Event
- December 3, 2016
- Report Date
- December 3, 2016
- Manufacturer
- DEXCOM INC.
- Product Code
- MDS
- UDI-DI
- 00386270000262
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4).
Description of Event or Problem · 1
DEXCOM WAS MADE AWARE ON (B)(6) 2016, THAT ON (B)(6) 2016, THE RECEIVER DISPLAYED AN ERROR 121. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. THE RECEIVER WAS RETURNED FOR EVALUATION. EXTERNAL VISUAL INSPECTION FOUND NO OBSERVATIONS RELATED TO THE CUSTOMER COMPLAINT. THE RECEIVER DATA LOG WAS DOWNLOADED AND REVIEWED AND FIRMWARE ERRORS WERE OBSERVED. THE REPORTED EVENT OF THE RECEIVER DISPLAYING ERROR CODE FAILURES WAS CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 847751 | DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM | CONTINUOUS GLUCOSE MONITOR | MDS | DEXCOM INC. | MT22719-PNK | 5213159 | 00386270000262 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 4 YR |