Report Number
3004753838-2016-81714
Event Type
Malfunction
Date Received
December 22, 2016
Date of Event
December 3, 2016
Report Date
December 3, 2016
Manufacturer
DEXCOM INC.
Product Code
MDS
UDI-DI
00386270000262
PMA / PMN Number
P120005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

DEXCOM WAS MADE AWARE ON (B)(6) 2016, THAT ON (B)(6) 2016, THE RECEIVER DISPLAYED AN ERROR 121. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. THE RECEIVER WAS RETURNED FOR EVALUATION. EXTERNAL VISUAL INSPECTION FOUND NO OBSERVATIONS RELATED TO THE CUSTOMER COMPLAINT. THE RECEIVER DATA LOG WAS DOWNLOADED AND REVIEWED AND FIRMWARE ERRORS WERE OBSERVED. THE REPORTED EVENT OF THE RECEIVER DISPLAYING ERROR CODE FAILURES WAS CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
847751 DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM CONTINUOUS GLUCOSE MONITOR MDS DEXCOM INC. MT22719-PNK 5213159 00386270000262

Patients

Seq Age Sex Outcome Treatment
1 4 YR