FDA Adverse Event
Malfunction
Summary report: N
S8-3T MICRO TEE TRANSDUCER
MDR report key: 6198934
·
Received December 22, 2016
Report
- Report Number
- 3019216-2016-00046
- Event Type
- Malfunction
- Date Received
- December 22, 2016
- Report Date
- December 8, 2016
- Manufacturer
- PHILIPS ULTRASOUND, INC
- Product Code
- ITX
- PMA / PMN Number
- K030455
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
EVALUATION OF THE TRANSDUCER WILL BE INCLUDED IN A FOLLOW UP REPORT UPON ITS RETURN AND INVESTIGATION COMPLETION.
Additional Manufacturer Narrative · 1
EVALUATION OF THE TRANSDUCER CONFIRMED OIL SEPARATION AND DAMAGE TO THE WINDOW. A SIGNIFICANT BUILDUP OF DEBRIS ON THE TIP CAP, WINDOW, BENDING NECK, AND BEAD WAS ALSO NOTED. THE CAUSE OF THE EXTENSIVE DAMAGE TO THIS DEVICE IS INDICATIVE OF IMPROPER HANDLING AND MAINTENANCE.
Description of Event or Problem · 1
A CUSTOMER REPORTED AN S8-3T MODEL TRANSDUCER WAS NOT PRODUCING AN IMAGE. THERE WAS NO INJURY ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 849722 | S8-3T MICRO TEE TRANSDUCER | S8-3T MICRO TEE TRANSDUCER | ITX | PHILIPS ULTRASOUND, INC | B1GTVH |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |