FDA Adverse Event
Malfunction
Summary report: N
980 VENTILATOR
MDR report key: 6197178
·
Received December 21, 2016
Report
- Report Number
- 8020893-2016-03582
- Event Type
- Malfunction
- Date Received
- December 21, 2016
- Report Date
- November 28, 2016
- Manufacturer
- COVIDIEN
- Product Code
- CBK
- PMA / PMN Number
- K131252
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE COVIDIEN SERVICE ENGINEER (SE) INSPECTED THE DEVICE AND VERIFIED THE REPORTED ISSUE. THE SE REPLACED THE OXYGEN SENSOR. THE SE PERFORMED EXTENDED SELF-TESTING ON THE DEVICE AND ALL TESTS PASSED.
Description of Event or Problem · 1
IT WAS REPORTED THAT, THE OXYGEN SENSOR ON A 980 VENTILATOR WOULD NOT CALIBRATE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME THE EVENT OCCURRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 847385 | 980 VENTILATOR | CONTINUOUS VENTILATOR | CBK | COVIDIEN | 980 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |