FDA Adverse Event Malfunction Summary report: N

MIDWEST TRADITION PB FIBER OPTIC

MDR report key: 6194359 · Received December 21, 2016

Report

Report Number
1419322-2016-00320
Event Type
Malfunction
Date Received
December 21, 2016
Report Date
December 21, 2016
Manufacturer
DENTSPLY PROFESSIONAL
Product Code
EFB
PMA / PMN Number
K963050
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

THOUGH NO MEDICAL/SURGICAL INTERVENTION WAS REQUIRED TO PRECLUDE A SERIOUS INJURY IN THIS EVENT, THERE HAVE BEEN PREVIOUSLY REPORTED EVENTS INVOLVING A SIMILAR DEVICE THAT RESULTED IN THE NEED FOR MEDICAL/SURGICAL INTERVENTION TO PRECLUDE PERMANENT DAMAGE TO A BODY STRUCTURE OR PERMANENT IMPAIRMENT OF A BODY FUNCTION. THEREFORE, THIS EVENT MEETS THE CRITERIA FOR REPORTABILITY PER 21 CFR PART 803. THE INVESTIGATION COULD NOT SUPPORT THE REPORTED COMPLAINT. THE RETURNED HANDPIECE WAS TESTED BY MANUFACTURING PERSONNEL AND DID MET ALL SPECIFICATIONS CRITERIA. QUALITY PERSONNEL THEN INVESTIGATED THE HANDPIECE. THE HANDPIECE EXHIBITED MAXIMUM TEMPERATURE OF 20.9°C DURING FREE RUN TESTING AND 15.8°C DURING LOAD TESTING. PER ISO 13732-1, THESE TEMPERATURE LEVELS DO NOT REACH THE BURN THRESHOLD REGARDLESS OF THE CONTACT PERIOD. MICROSCOPIC EVALUATION REVEALED DEBRIS WITHIN THE CAP AND HEAD CAVITIES AND CAP END BEARING OUTER RACE. MICROSCOPIC EVALUATION REVEALED SIGNIFICANT WEAR ON THE INSIDE SURFACE OF THE CAP BUTTON THAT FACES THE PUSHER AND ON THE PUSHER ITSELF. THIS INDICATES SEVERE INTERACTION BETWEEN THESE TWO MECHANISMS AT HIGH ROTATIONAL SPEEDS WHICH MOST LIKELY CREATES HEAT NOTED IN THE COMPLAINT.

Description of Event or Problem · 1

IN THIS EVENT A DOCTOR REPORTED THAT A TRADITION HANDPIECE OVERHEATED. THERE WAS NO INJURY OR INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
842772 MIDWEST TRADITION PB FIBER OPTIC HANDPIECE, AIR-POWERED, DENTAL EFB DENTSPLY PROFESSIONAL NA NA

Patients

Seq Age Sex Outcome Treatment
1