R SERIES DEFIBRILLATOR
MDR report key: 6193507
·
Received December 21, 2016
Report
- Report Number
- 1220908-2016-03196
- Event Type
- Malfunction
- Date Received
- December 21, 2016
- Report Date
- December 6, 2016
- Manufacturer
- ZOLL MEDICAL CORPORATION
- Product Code
- MKJ
- PMA / PMN Number
- K060559
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
ZOLL MEDICAL CORPORATION EVALUATED THE DEVICE AND THE REPORTED MALFUNCTION WAS NOT REPLICATED OR CONFIRMED. THE DEVICE WAS PUT THROUGH EXTENSIVE TESTING WITHOUT DUPLICATING THE MALFUNCTION. THE DEVICE'S COLOR LCD DISPLAY CABLE WAS REPLACED AS A PRECAUTION. THE DEVICE WAS RECERTIFIED AND RETURNED TO THE CUSTOMER. NO TREND IS ASSOCIATED WITH REPORTS OF THIS TYPE.
Description of Event or Problem · 1
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE INTERMITTENTLY POWERED UP WITH A DISTORTED DISPLAY WITH MULTIPLE COLORS IN THE BACKGROUND AND NO CLINICAL INFORMATION COULD BE SEEN. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 846932 | R SERIES DEFIBRILLATOR | DEFIBRILLATOR/PACEMAKER | MKJ | ZOLL MEDICAL CORPORATION | R SERIES | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |