TI MATRIXNEURO SCREW SELF-DRILLING 4MM
Report
- Report Number
- 1719045-2016-10936
- Event Type
- Malfunction
- Date Received
- December 21, 2016
- Date of Event
- November 29, 2016
- Report Date
- November 29, 2016
- Manufacturer
- SYNTHES MONUMENT
- Product Code
- JEY
- UDI-DI
- 20887587018345
- PMA / PMN Number
- K042365
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- OTHER
Narratives
DEVICE BROKE DURING INSERTION; DEVICE WAS NOT IMPLANTED/EXPLANTED. COMPLAINANT PART IS NOT EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION. (B)(6). STERILE KIT PART 145.321S, LOT L092044: MANUFACTURING LOCATION: (B)(4). SUPPLIER: (B)(4). MANUFACTURING DATE: AUGUST 11, 2016. EXPIRY DATE: AUGUST 01, 2026. NON-STERILE SCREW PART 04.503.104.20, LOT H110805: MANUFACTURING LOCATION: (B)(4). MANUFACTURING DATE: MAY 24, 2016. THIS COMPLAINT IS ASSESSED AS NOT RELATED TO STERILIZATION. NO NON-CONFORMANCE REPORTS WERE GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORD(S) SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THE PRODUCT THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
DEVICE REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT DURING IMPLANTATION THE HEAD OF A SCREW FROM A MATRIX NEURO STERILE KIT BROKE. THE SURGERY WAS NOT PROLONGED. THERE WAS NO PATIENT HARM AND ALL BROKEN OFF PIECES WERE REMOVED. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT. THE SURGERY WAS SUCCESSFULLY COMPLETED. THIS IS REPORT 1 OF 1 FOR COM-(B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 842881 | TI MATRIXNEURO SCREW SELF-DRILLING 4MM | BONE PLATE | JEY | SYNTHES MONUMENT | H110805 | 20887587018345 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |