FDA Adverse Event Injury Summary report: N

UNKNOWN

MDR report key: 6192229 · Received December 21, 2016

Report

Report Number
3002808486-2016-01524
Event Type
Injury
Date Received
December 21, 2016
Report Date
August 22, 2024
Manufacturer
WILLIAM COOK EUROPE
Product Code
DTK
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
ATTORNEY
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

CORRECTION: A2, A4. ADDITIONAL INFORMATION: B5, B6, B7, D6B, H6. INVESTIGATION: INVESTIGATION IS REOPENED DUE TO ADDITIONAL INFORMATION PROVIDED. THE REPORTED ALLEGATIONS HAVE BEEN FURTHER INVESTIGATED BASED ON THE INFORMATION PROVIDED TO DATE. THE FOLLOWING ALLEGATIONS HAVE BEEN INVESTIGATED: ORGAN (ORG)/VENA CAVA (VC) PERFORATION, MIGRATION, COMPLEX RETRIEVAL, BACK PAIN DISCOMFORT, LIMITED PHYSICAL ACTIVITY. FILTER INTERACTS WITH IVC WALL, E.G. PENETRATION/PERFORATION/EMBEDMENT. THIS MAY BE EITHER SYMPTOMATIC OR ASYMPTOMATIC. POTENTIAL CAUSES MAY INCLUDE IMPROPER DEPLOYMENT; AND (OR) EXCESSIVE FORCE OR MANIPULATIONS NEAR AN IN-SITU FILTER (E.G., A SURGICAL OR ENDOVASCULAR PROCEDURE IN THE VICINITY OF A FILTER). POTENTIAL ADVERSE EVENTS THAT MAY OCCUR INCLUDE, BUT ARE NOT LIMITED TO, THE FOLLOWING: TRAUMA TO ADJACENT STRUCTURES, VASCULAR TRAUMA. PHYSICIAN PRACTICE GUIDELINES AND PUBLISHED GUIDANCE FROM REGULATORY AGENCIES RECOMMEND THAT PATIENTS WITH INDWELLING FILTERS UNDERGO ROUTINE FOLLOW-UP. THE RISKS/BENEFITS OF FILTER RETRIEVAL SHOULD BE CONSIDERED FOR EACH PATIENT DURING FOLLOW-UP. ONCE PROTECTION FROM PE IS NO LONGER NECESSARY, FILTER RETRIEVAL SHOULD BE CONSIDERED. FILTER RETRIEVAL SHOULD BE ATTEMPTED WHEN FEASIBLE AND CLINICALLY INDICATED. FILTER RETRIEVAL IS A PATIENT-SPECIFIC, CLINICALLY COMPLEX DECISION; THE DECISION TO REMOVE A FILTER SHOULD BE BASED ON EACH PATIENT¿S INDIVIDUAL RISK/BENEFIT PROFILE (E.G., A PATIENT¿S CONTINUED NEED FOR PROTECTION FROM PE COMPARED TO THEIR EXPERIENCE WITH AND (OR) ONGOING RISK OF EXPERIENCING FILTER-RELATED COMPLICATIONS). THE FILTER IS DESIGNED TO BE RETRIEVED WITH THE GÜNTHER TULIP VENA CAVA FILTER RETRIEVAL SET. IT MAY ALSO BE RETRIEVED WITH THE CLOVERSNARE® VASCULAR RETRIEVER. COOK HAS NOT PERFORMED TESTING TO EVALUATE THE SAFETY OR EFFECTIVENESS OF FILTER RETRIEVAL USING OTHER RETRIEVAL SYSTEMS OR TECHNIQUES. THE PUBLISHED CLINICAL LITERATURE INCLUDES DESCRIPTIONS OF ALTERNATIVE TECHNIQUES FOR FILTER RETRIEVAL; USE OF THESE TECHNIQUES VARIES ACCORDING TO PHYSICIAN EXPERIENCE, PATIENT ANATOMY, AND FILTER POSITION. THE SAFETY OR EFFECTIVENESS OF THESE ALTERNATIVE RETRIEVAL TECHNIQUES HAS NOT BEEN ESTABLISHED. SPECIFIC FOR ¿EMBEDDED¿ A FILTER THAT IS EMBEDDED IN THE WALL OF THE IVC MAY BE DIFFICULT TO RETRIEVE. FOR ALL RETRIEVABLE IVC FILTERS, RETRIEVAL BECOMES MORE CHALLENGING WITH TIME, AND THIS IS COMMONLY DUE TO ENCAPSULATION OF THE FILTER LEGS OR HOOK (IN A TILTED FILTER) BY TISSUE INGROWTH. FILTER OR FILTER FRAGMENT MIGRATION AND (OR) EMBOLIZATION (E.G., MOVEMENT TO THE HEART OR LUNGS) HAS BEEN REPORTED. FILTER OR FILTER FRAGMENT MOVEMENT HAS OCCURRED IN BOTH THE CRANIAL AND CAUDAL DIRECTION AND MAY BE EITHER SYMPTOMATIC OR ASYMPTOMATIC. POTENTIAL CAUSES MAY INCLUDE FILTER PLACEMENT IN IVCS WITH DIAMETERS SMALLER OR LARGER THAN THOSE SPECIFIED IN THESE INSTRUCTIONS FOR USE; IMPROPER DEPLOYMENT; DEPLOYMENT INTO THROMBUS; DISLODGEMENT DUE TO LARGE THROMBUS BURDENS; AND (OR) EXCESSIVE FORCE OR MANIPULATIONS NEAR AN IN SITU FILTER (E.G., A SURGICAL OR ENDOVASCULAR PROCEDURE IN THE VICINITY OF A FILTER). POTENTIAL ADVERSE EVENTS THAT MAY OCCUR INCLUDE, BUT ARE NOT LIMITED TO, THE FOLLOWING: FILTER MIGRATION, TRAUMA TO ADJACENT STRUCTURES. UNKNOWN IF THE REPORTED BACK PAIN, DISCOMFORT, AND LIMITED PHYSICAL ACTIVITY ARE DIRECTLY RELATED TO THE FILTER AND UNABLE TO IDENTIFY A CORRESPONDING FAILURE MODE AT THIS POINT IN TIME. CATALOG AND LOT NUMBERS ARE UNKNOWN, HOWEVER, THE ALLEGED TULIP IS MANUFACTURED AND INSPECTED ACCORDING TO CONTROLS. NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS AND NOTHING INDICATES THAT THE FILTER DID NOT PERFORM AS INTENDED, E.G. INTENDED FOR THE PREVENTION OF RECURRENT PULMONARY EMBOLISM (PE) VIA PLACEMENT IN THE VENA CAVA. COOK WILL REOPEN ITS INVESTIGATION IF FURTHER INFORMATION IS RECEIVED WARRANTING SUPPLEMENTATION IN ACCORDANCE WITH 21 C.F.R. 803.56. THIS REPORT INCLUDES INFORMATION KNOWN AT THIS TIME. A FOLLOW-UP REPORT WILL BE SUBMITTED SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803 AND IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT CONTAINED HEREIN IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED, THAT A DEATH OR SERIOUS INJURY OCCURRED, NOR THAT ANY COOK DEVICE CAUSED, CONTRIBUTED TO, OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.

Description of Event or Problem · 0

ADDITIONAL INFORMATION WAS RECEIVED AS FOLLOWS: REASON FOR RETRIEVAL - IN ADDITION TO ALLEGATIONS PREVIOUSLY REPORTED, PATIENT ALLEGES FILTER MIGRATION, VENA CAVA PERFORATION, AND LEGS OF FILTER ERODED THROUGH IVC AND INTO ADJACENT LUMBAR VERTEBRAL BODY. PATIENT NOTES AND FURTHER ALLEGES EXPERIENCING PAIN, DISCOMFORT, LIMITED PHYSICAL ACTIVITY-ESPECIALLY WALKING. PER THE RETRIEVAL REPORT (SUCCESSFUL): "MORE RECENTLY, A CT EXAMINATION SHOWED ONE OF THE LEGS OF IVC FILTER ERODED INTO THE ADJACENT LUMBAR VERTEBRAL BODY. PATIENT CONTINUES OF CONSTANT LOWER BACK PAIN AND FELT IT MAY BE RELATED". "IMPRESSION: 1. IVC VENOGRAM SHOWING NO EVIDENCE OF THROMBUS WITHIN THE FILTER WITH THE CRANIAL TIP EMBEDDED IN THE IVC WALL. 2. UNEVENTFUL ULTRASOUND AND FLUOROSCOPY GUIDED COMPLEX RETRIEVAL OF A PREVIOUSLY PLACED INFRARENAL GUNTHER-TULIP IVC FILTER".

Additional Manufacturer Narrative · 1

(B)(4). CATALOG#: UNKNOWN BUT REFERRED TO AS A COOK GÜNTHER TULIP FILTER. SINCE CATALOG# IS UNKNOWN THE 510(K) COULD BE EITHER K090140, K112119 OR K121057 (B)(4). IT HAS NOT BEEN POSSIBLE TO INVESTIGATE OR EVALUATE THIS ALLEGED EVENT BASED ON THE LIMITED INFORMATION PROVIDED TO DATE. COOK WILL REOPEN ITS INVESTIGATION IF FURTHER INFORMATION IS RECEIVING WARRANTING SUPPLEMENTATION IN ACCORDANCE WITH 21 C.F.R. 803.56.

Additional Manufacturer Narrative · 1

EXEMPTION NUMBER E LOT. (B)(4). NAME AND ADDRESS FOR IMPORTER SITE: (B)(4). 510(K) COULD BE ANY OF THE FOLLOWING: K112119; K121629 THE EVENT IS CURRENTLY UNDER INVESTIGATION. A SUPPLEMENTAL REPORT WILL BE PROVIDED UPON CONCLUSION.

Additional Manufacturer Narrative · 1

(B)(4). CATALOG #: UNKNOWN BUT REFERRED TO AS A COOK GÜNTHER TULIP FILTER. EXPIRATION DATE: UNKNOWN AS LOT # IS UNKNOWN. SINCE CATALOG # IS UNKNOWN THE 510(K) COULD BE EITHER K090140, K112119 OR K121057. MFR DATE UNKNOWN AS LOT # IS UNKNOWN. INVESTIGATION IS STILL IN PROGRESS.

Additional Manufacturer Narrative · 1

MANUFACTURER REFERENCE # (B)(4). EXEMPTION NUMBER E2016032. (B)(4). IT HAS NOT BEEN POSSIBLE TO FURTHER INVESTIGATE OR EVALUATE THIS ALLEGED EVENT BASED ON THE LIMITED INFORMATION PROVIDED TO DATE VIA THE OPERATIVE NOTE STATING 'GUNTHER TULIP - UNABLE TO RETRIEVE; BLOOD CLOTS IN LUNGS; THROMBUS IN APEX; EMBEDDED; FEAR'. COOK WILL REOPEN ITS INVESTIGATION IF FURTHER INFORMATION IS RECEIVED WARRANTING SUPPLEMENTATION IN ACCORDANCE WITH 21 C.F.R. 803.56. A FILTER THAT IS EMBEDDED IN THE WALL OF THE IVC MAY BE DIFFICULT TO RETRIEVE. FILTER RETRIEVAL IS OCCASIONALLY DIFFICULT. THIS IS WELL-KNOWN FROM PUBLISHED SCIENTIFIC LITERATURE WHERE FILTER RETRIEVALS ARE REFERRED TO AS SIMPLE VS. COMPLEX. SEVERAL CASE REPORTS PUBLISHED IN SCIENTIFIC LITERATURE DESCRIBE COMPLEX CASES WITH SUCCESSFUL ENDOVASCULAR FILTER RETRIEVALS USING ADDITIONAL, ADVANCED TECHNIQUES. PE IS A KNOWN RISK IN RELATION TO FILTER IMPLANT REPORTED IN THE PUBLISHED SCIENTIFIC LITERATURE. ALSO, IT IS REPORTED THAT THE PULMONARY EMBOLISM IN SOME CASES MAY ORIGINATE FROM UPPER EXTREMITIES INSTEAD OF LOWER EXTREMITY VEINS. WITH ALL FILTERS, THERE IS SOME RISK OF FURTHER PULMONARY EMBOLISM. IVC THROMBOTIC OCCLUSION AS AN OUTCOME FOR COOK IVC FILTERS IS ADDRESSED IN THE PUBLISHED SCIENTIFIC LITERATURE. IVC THROMBOTIC OCCLUSION IS DEFINED AS THE PRESENCE OF AN OCCLUDING THROMBUS IN THE IVC AFTER FILTER INSERTION AND DOCUMENTED BY ULTRASOUND (US), CT, MR IMAGING OR VENOGRAPHY; THIS MAY BE SYMPTOMATIC OR ASYMPTOMATIC. UNKNOWN IF THE REPORTED FEAR IS DIRECTLY RELATED TO THE FILTER AND UNABLE TO IDENTIFY CORRESPONDING FAILURE MODE(S) AT THIS TIME. NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS AND NOTHING INDICATES THAT THE FILTER DID NOT PERFORM AS INTENDED, E.G. INTENDED FOR THE PREVENTION OF RECURRENT PULMONARY EMBOLISM (PE) VIA PLACEMENT IN THE VENA CAVA.

Description of Event or Problem · 1

DESCRIPTION ACCORDING TO SHORT FORM COMPLAINT FILED: IT IS ALLEGED THAT "[PT] RECEIVED A COOK GUNTHER TULIP FILTER ON (B)(6) 2014". PATIENT OUTCOME: IT IS ALLEGED THAT [PT] WAS INJURED WITHOUT FURTHER EXPLANATION. HOSPITAL AND MEDICAL RECORDS HAVE BEEN REQUESTED BUT NOT YET PROVIDED.

Description of Event or Problem · 1

THIS ADDITIONAL INFORMATION RECEIVED ON 25MAY2017AS FOLLOWS: [PT] ALLEGEDLY RECEIVED AN IMPLANT ON (B)(6) 2014 DUE TO A PULMONARY EMBOLISM. [PT] IS ALLEGING THE DEVICE IS UNABLE TO BE RETRIEVED. A RETRIEVAL WAS ATTEMPTED ON (B)(6) 2014 BUT WAS UNSUCCESSFUL DUE TO THE FILTER LEGS BEING EMBEDDED IN THE IVC. [PT] FURTHER ALLEGES TO HAVE BLOOD CLOTS IN LUNGS, AND FEAR THAT THE DEVICE MAY BREAK LOSE AND LEAD TO DEATH.

Description of Event or Problem · 1

DESCRIPTION ACCORDING TO SHORT FORM COMPLAINT FILED: IT IS ALLEGED THAT "[PT] RECEIVED A COOK GUNTHER TULIP FILTER ON (B)(6) 2014." PATIENT OUTCOME: IT IS ALLEGED THAT [PT] WAS INJURED WITHOUT FURTHER EXPLANATION. HOSPITAL AND MEDICAL RECORDS HAVE BEEN REQUESTED BUT NOT YET PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
843624 UNKNOWN DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR DTK WILLIAM COOK EUROPE

Patients

Seq Age Sex Outcome Treatment
1 18 YR Female Life Threatening