FDA Adverse Event Other Summary report: N

SPYGLASS ANGIOGRAPHIC CATH. 4F ANGLED PIGTAIL

MDR report key: 619178 · Received June 22, 2005

Report

Report Number
2182269-2005-00114
Event Type
Other
Date Received
June 22, 2005
Date of Event
May 20, 2005
Report Date
June 22, 2005
Manufacturer
ST. JUDE MEDICAL
Product Code
DQO
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

IT WAS REPORTED DURING THE CARDIAC CATHETERIZATION PROCEDURE, THE TIP OF THE 4F SPYGLASS PIGTAIL CATHETER DETACHED WHILE IN THE PT'S AORTA. THE RIGHT AND LEFT HEART CATHETERIZATION PROCEDURE WERE DONE TO EVALUATE THE PT'S MITRAL REGURGITATION, PULMONARY HYPERTENSION, AND ANGINA. THE PT ALSO HAD SEVERE PERIPHERAL VASCULAR DISEASE. THE RIGHT HEART CATHETERIZATION PROCEDURE WAS PERFORMED VIA THE RIGHT FEMORAL VEIN AND THE LEFT HEART CATHETERIZATION WAS PERFORMED VIA THE RIGHT BRACHIAL ARTERY. HEPARIN, DOSE UNK WAS GIVEN DURING THE PROCEDURE. THE 4F SPYGLASS PIGTAIL CATHETER WAS ADVANCED INTO THE PT WITH THE INTENT TO PERFORM VENTRICULOGRAPHY, WHEN THE TIP OF THE CATHETER SEPARATED AND EMBOLIZED TO THE ABDOMINAL AORTA. THE GUIDE WIRE WAS ADVANCED TO THE AREA AND A 6F MULTIPURPOSE GUIDING CATHETER WAS THEN ADVANCED TO THE DETACHED CATHETER TIP. A 7 MILLIMETER AMPLATZ GOOSE NECK MICROSNARE CATHETER WAS ADVANCED AND THE TIP AND WAS REMOVED BY THE CARDIOLOGIST. RESULTS OF THE ANGIOGRAM REVEALED THE ARTHEROSCLEROTIC IRREGULARITY AND 20-30% STENOSIS OF THE DOMINANT RIGHT CORONARY ARTERY AND AORTIC OCCLUSION BELOW THE SUPERIOR MESENTERIO ARTERY WITH RECONSTITUTION AT THE LEVEL OF THE COMMON FEMORAL ARTERIES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPYGLASS ANGIOGRAPHIC CATH. 4F ANGLED PIGTAIL SPYGLASS 4F ANGLED PIGTAIL DQO ST. JUDE MEDICAL NA 1074612

Patients

Seq Age Sex Outcome Treatment
1 37 YR Required Intervention