FDA Adverse Event Summary report: N

LIGACLIP MCA MULTIPLE CLIP APPLIER

MDR report key: 6190521 · Received December 20, 2016

Report

Report Number
3005075853-2016-07185
Date Received
December 20, 2016
Report Date
December 14, 2016
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDO
PMA / PMN Number
K820837
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE MSM20 DEVICE WAS RETURNED WITH A CLIP IN THE JAWS. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED AND IT FED AND FORMED THE REMAINING 9 CLIPS AS INTENDED. AS THE DEVICE WAS FOUND TO BE FULLY FUNCTIONAL, IT COULD NOT BE DETERMINED WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH HISTORY RECORD WAS REVIEWED AND NO DEFECTS, NCR¿S OR PROTOCOLS RELATED TO THE COMPLAINT, WERE FOUND DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4). THE LOT HISTORY RECORDS WERE REVIEWED AND THE MANUFACTURING CRITERIA WERE MET PRIOR TO THE RELEASE OF THIS LOT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE CLIP WAS NOT RELEASED. THE EXERTED PRESSURE HAS CUT THE VESSEL. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
840482 LIGACLIP MCA MULTIPLE CLIP APPLIER CLIP, IMPLANTABLE GDO ETHICON ENDO-SURGERY, LLC. NA N4LT42

Patients

Seq Age Sex Outcome Treatment
1 Other