LIGACLIP MCA MULTIPLE CLIP APPLIER
Report
- Report Number
- 3005075853-2016-07185
- Date Received
- December 20, 2016
- Report Date
- December 14, 2016
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDO
- PMA / PMN Number
- K820837
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE ANALYSIS RESULTS FOUND THAT THE MSM20 DEVICE WAS RETURNED WITH A CLIP IN THE JAWS. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED AND IT FED AND FORMED THE REMAINING 9 CLIPS AS INTENDED. AS THE DEVICE WAS FOUND TO BE FULLY FUNCTIONAL, IT COULD NOT BE DETERMINED WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH HISTORY RECORD WAS REVIEWED AND NO DEFECTS, NCR¿S OR PROTOCOLS RELATED TO THE COMPLAINT, WERE FOUND DURING THE MANUFACTURING PROCESS.
(B)(4). THE LOT HISTORY RECORDS WERE REVIEWED AND THE MANUFACTURING CRITERIA WERE MET PRIOR TO THE RELEASE OF THIS LOT.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE CLIP WAS NOT RELEASED. THE EXERTED PRESSURE HAS CUT THE VESSEL. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 840482 | LIGACLIP MCA MULTIPLE CLIP APPLIER | CLIP, IMPLANTABLE | GDO | ETHICON ENDO-SURGERY, LLC. | NA | N4LT42 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |