FDA Adverse Event Injury Summary report: N

VAC-PAC

MDR report key: 618984 · Received June 9, 2005

Report

Report Number
618984
Event Type
Injury
Date Received
June 9, 2005
Date of Event
April 8, 2005
Report Date
June 9, 2005
Manufacturer
OLYMPIC MEDICAL CORP.
Product Code
CCX
Adverse Event
Yes
Report Source
User Facility report
Reporter Location
AZ, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

LARGE BODY MASS PATIENT WAS HAVING AN OPEN REDUCTION WITH INTERNAL FIXATION ON THE RIGHT WRIST. THE PATIENT WAS POSITIONED IN THE LEFT LATERAL POSITION ON THE BEAN BAG AND THE AIR WAS SUCTIONED OUT OF THE BAG. A BLANKET WAS PLACED AT THE BASE OF THE PAD BUT NOTHING WAS BETWEEN THE PATIENT AND THE PAD. A PILLOW WAS ALSO PLACED BETWEEN THE PATIENT'S LEGS AND GEL PADS UNDER THE PATIENT'S LEG WHICH WAS IN CONTACT WITH THE TABLE. PART WAY THROUGH THE PROCEDURE, THE PATIENT WAS MOVED ON THE TABLE TO FACILITATE SURGERY WITHOUT UNDRAPING OR RECHECKING THE POSITION. AT THE END OF CASE, A RED SPOT NOTED ON PATIENT'S LEFT THIGH AND THE PATIENT COMPLAINED OF PAIN AT THE SITE. DIAGNOSED WITH COMPARTMENT SYNDROME THE NEXT MORNING SUSPECTED TO BE FROM THE PATIENT'S TOP LEG WEIGHTED ON THE LOWER LEG ALONG WITH THE PRESSURE FROM THE EDGE OF THE AIR BAG. FASCIOTOMIES WERE PERFORMED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VAC-PAC BEAN BAG POSITIONING DEVICE CCX OLYMPIC MEDICAL CORP. * *

Patients

Seq Age Sex Outcome Treatment
1 24 YR Hospitalization| R