HEARTSINE SAMARITAN 300P AND PAD PAK
Report
- Report Number
- 3004123209-2016-01871
- Event Type
- Malfunction
- Date Received
- December 20, 2016
- Date of Event
- November 28, 2016
- Report Date
- December 19, 2016
- Manufacturer
- HEARTSINE TECHNOLOGIES LTD
- Product Code
- MKJ
- PMA / PMN Number
- K041067
- Removal / Correction Number
- Z-0124-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- PATIENT
Narratives
EXEMPTION NUMBER E2015058. HEARTSINE TECHNOLOGIES LTD (MANUFACTURER) IS SUBMITTING THE REPORT ON BEHALF OF HEARTSINE TECHNOLOGIES LLC (IMPORTER). HEARTSINE TECHNOLOGIES LTD (MANUFACTURER) IS SUBMITTING THE REPORT ON BEHALF OF (B)(4). ROUTINE TESTING CONFIRMED THAT THE PAD-PAK WAS FIRST WAS INSTALLED BY THE CUSTOMER ON THE (B)(6) 2009. THE DEVICE RECORDS MANUAL POWER UPS OF TEN MINUTES DURATION BETWEEN THE (B)(6) 2016. INVESTIGATION FOUND THE FAULT CAN BE ATTRIBUTED TO MEMBRANE FAILURE. THE PAD-PAK, WHICH CONTAINS THE ELECTRODES AND BATTERIES, IS LABELLED FOR SINGLE USE BUT THE SAMARITAN PAD 300 AND 300P DEVICES ARE FOR MULTI-USE.
EXEMPTION NUMBER E2015058 HEARTSINE TECHNOLOGIES LTD (MANUFACTURER) IS SUBMITTING THE REPORT ON BEHALF OF HEARTSINE TECHNOLOGIES LLC (IMPORTER). H3 OTHER TEXT : NOT YET RETURNED TO MANUFACTURER
THERE WAS NO PATIENT INVOLVED IN THIS EVENT. DEVICE SWITCHING ON AUTOMATICALLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 842279 | HEARTSINE SAMARITAN 300P AND PAD PAK | AUTOMATED EXTERNAL DEFIBRILLATOR | MKJ | HEARTSINE TECHNOLOGIES LTD | SAM | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |