FDA Adverse Event Malfunction Summary report: N

INOMAX DSIR (DELIVERY SYSTEM)

MDR report key: 6189137 · Received December 20, 2016

Report

Report Number
3004531588-2016-00083
Event Type
Malfunction
Date Received
December 20, 2016
Report Date
December 20, 2016
Manufacturer
MALLINCKRODT MANUFACTURING LLC
Product Code
MRN
PMA / PMN Number
K061901
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE ISSUE WITH INOMAX DSIR (B)(4) WAS CREATED AS (B)(4). THE DEVICE INVESTIGATION WAS COMPLETED ON 08-DEC-2016. THE REGIONAL SERVICE CENTER (RSC) SERVICE LOG REVIEW REVEALED A DELIVERY FAILURE ALARM DUE TO INTERPROCESSOR WATCHDOG TIMED OUT BUT WAS ABLE TO CONFIRM IT IN THE SERVICE LOG. THE 50018 MAIN BOARD WAS REPLACED AS PRECAUTION DUE TO THE SERVICE LOG FINDINGS. A FULL FUNCTIONAL TEST WAS PERFORMED AND THE DEVICE OPERATED ACCORDING TO SPECIFICATIONS SO IT WAS RETURNED TO THE DEVICE SERVICE POOL. THE ROOT CAUSE FOR THIS REPORT WAS DELIVERY FAILURE; APPARENT IPL FAILURE DUE TO MAIN BOARD FAILURE. THIS CONDITION WILL BE TRACKED AND TRENDED UNDER THE COMPANY'S QUALITY SYSTEM. THIS CASE DID NOT RESULT IN AN ADVERSE EVENT/SERIOUS ADVERSE EVENT, HOWEVER, IT IS BEING SUBMITTED TO REGULATORY AUTHORITIES BECAUSE A SIMILAR FAILURE OCCURRED IN THE PAST WHICH RESULTED IN A SERIOUS ADVERSE EVENT (MDR 3004531588-2016-00070).

Description of Event or Problem · 1

ON 04-NOV-2016, A RESPIRATORY THERAPIST (RT) FROM THE UNITED STATES CALLED MALLINCKRODT CUSTOMER CARE TO REPORT A DEVICE ISSUE WITH INOMAX DSIR (B)(4), UNRELATED TO THE REPORTABLE MALFUNCTION. DURING INTERNAL REVIEW OF THE SERVICE LOG, AN INTERNAL EMPLOYEE IDENTIFIED A DELIVERY FAILURE ALARM DUE TO INTERPROCESSOR WATCHDOG TIMED OUT. INOMAX DSIR (B)(4) WAS REMOVED FROM SERVICE AND RETURNED TO THE COMPANY FOR SERVICE EVALUATION. THIS IS BEING REPORTED AS IT IS CONSIDERED A REPORTABLE MALFUNCTION BASED ON THE COMPLETED INVESTIGATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
840525 INOMAX DSIR (DELIVERY SYSTEM) APPARATUS MRN MALLINCKRODT MANUFACTURING LLC 10007

Patients

Seq Age Sex Outcome Treatment
1