FDA Adverse Event Malfunction Summary report: N

OPTIFLUX 180NRE DIALYZER FINISHED ASSY.

MDR report key: 6186431 · Received December 19, 2016

Report

Report Number
1713747-2016-00577
Event Type
Malfunction
Date Received
December 19, 2016
Date of Event
November 21, 2016
Report Date
December 19, 2016
Manufacturer
OGDEN MANUFACTURING PLANT
Product Code
FJI
PMA / PMN Number
K002761
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED COMPLAINT IS NOT CONFIRMED AS THE COMPLAINT DEVICE WAS NOT AVAILABLE FOR MANUFACTURER EVALUATION. AS SUCH, A DEFINITIVE CONCLUSION REGARDING THE COMPLAINT INCIDENT CANNOT BE REACHED WITHOUT A PHYSICAL EXAMINATION OF THE COMPLAINT DEVICE. A RECORDS REVIEW WAS PERFORMED ON THE REPORTED LOT. AN INVESTIGATION OF THE DEVICE MANUFACTURING RECORDS WAS CONDUCTED BY THE MANUFACTURER. THERE WERE NO DEVIATIONS OR NON-CONFORMANCES IDENTIFIED DURING THE MANUFACTURING PROCESS WHICH COULD BE ASSOCIATED WITH THE REPORTED EVENT. IN ADDITION, THE BATCH RECORD REVIEW CONFIRMED THE LABELING, MATERIAL, AND PROCESS CONTROLS WERE WITHIN SPECIFICATION. THE LOT PASSED ALL RELEASE CRITERIA. REVIEW OF THE BATCH PRODUCTION RECORD (BPR) DID NOT REVEAL A PROBABLE CAUSE FOR THE CUSTOMER COMPLAINT.

Description of Event or Problem · 1

A USER FACILITY REPORTED THAT A BLOOD LEAK OCCURRED DURING THE PATIENT'S HEMODIALYSIS (HD) TREATMENT. THE BLOOD LEAK WAS NOTED AS BEING AN EXTERNAL DIALYZER LEAK. THE MACHINE DID NOT ALARM AS IT WAS NOT INTENDED TO DO SO. BLOOD WAS VISUALLY OBSERVED LEAKING FROM UNDER THE HEADER END CAP OF THE DIALYZER. THEREFORE, BLOOD TEST STRIPS WERE NOT USED TO CONFIRM THE PRESENCE OF BLOOD. NO DIALYZER DAMAGE WAS VISIBLE. THE PATIENT¿S ESTIMATED BLOOD LOSS (EBL) WAS NOTED AS BEING APPROXIMATELY 10ML. NO PATIENT ADVERSE EFFECTS WERE EXPERIENCED AND NO MEDICAL INTERVENTION WAS REQUIRED AS A RESULT OF THIS EVENT. THE PATIENT COMPLETED TREATMENT WITH A NEW SET-UP ON THE SAME MACHINE. THE COMPLAINT DEVICE IS NOT AVAILABLE TO BE RETURNED TO THE MANUFACTURER FOR EVALUATION AS IT WAS DISCARDED BY THE USER FACILITY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
838581 OPTIFLUX 180NRE DIALYZER FINISHED ASSY. DIALYZER, CAPILLARY, HOLLOW FIBER FJI OGDEN MANUFACTURING PLANT 16CU05002

Patients

Seq Age Sex Outcome Treatment
1 61 YR