ENVISION BLOWER KIT
Report
- Report Number
- 1824206-2016-00532
- Event Type
- Injury
- Date Received
- December 19, 2016
- Date of Event
- November 21, 2016
- Report Date
- November 21, 2016
- Manufacturer
- HILL-ROM BATESVILLE
- Product Code
- IOQ
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- NURSE
Narratives
THE PRESSURE WOUND WAS TREATED WITH DAKINS SOLUTION AND ABDOMINAL PADS. DEVELOPMENT OF PRESSURE ULCERS IS MULTIFACTORIAL AND CANNOT BE ONLY ATTRIBUTED TO PERFORMANCE OF THE SURFACE. RISK FACTORS INCLUDE PROTEIN-CALORIE MALNUTRITION, MICROCLIMATE (SKIN WETNESS CAUSED BY SWEATING OR INCONTINENCE), DISEASES THAT REDUCE BLOOD FLOW TO THE SKIN, SUCH AS ARTERIOSCLEROSIS, OR DISEASES THAT REDUCE THE SENSATION IN THE SKIN, SUCH AS PARALYSIS OR NEUROPATHY. POSITION CHANGES ARE KEY TO PRESSURE SORE PREVENTION AND TREATMENT. THESE CHANGES NEED TO BE FREQUENT, REPOSITIONING NEEDS TO AVOID STRESS ON THE SKIN, AND BODY POSITIONS NEED TO MINIMIZE THE RISK OF PRESSURE ON VULNERABLE AREAS. THE BLOWER CONNECTION WAS NOT PROPERLY ATTACHED CAUSING THE BLOWER TO NOT FUNCTION AS DESIGNED. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BLOWER IN AUGUST 2016. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BLOWER. THE TECHNICIAN PROPERLY CONNECTED THE BLOWER TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE PATIENT DEVELOPED A PRESSURE ULCER ON HIS SACRUM THAT PROGRESSED TO STAGE 4. THE BLOWER WAS LOCATED AT THE ACCOUNT. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 837310 | ENVISION BLOWER KIT | BED FLOTATION THERAPY POWERED | IOQ | HILL-ROM BATESVILLE | P136026 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |