FDA Adverse Event Injury Summary report: N

HT70 PLUS VENTILATOR

MDR report key: 6186285 · Received December 19, 2016

Report

Report Number
2023050-2016-00754
Event Type
Injury
Date Received
December 19, 2016
Report Date
October 4, 2018
Manufacturer
NEWPORT MEDICAL INSTRUMENTS, INC
Product Code
CBK
PMA / PMN Number
K111146
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4)

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

(B)(4). A COVIDIEN FIELD SERVICE ENGINEER (FSE) INSPECTED THE DEVICE AND VERIFIED THE REPORTED CONDITION. THEY REPLACED THE PUMP AND THE MIXER GASKET. THE FSE PERFORMED EXTENDED SELF-TESTING ON THE DEVICE AND ALL TESTS PASSED.

Additional Manufacturer Narrative · 1

UNIQUE IDENTIFIER (UDI) NUMBER ADDED. CORRECTION TO CONTACT INFORMATION. CORRECTION TO EVALUATION CODES RESULT AND CONCLUSION. DEVICE EVALUATION SUMMARY: THE PUMP/MANIFOLD ASSEMBLY WAS RETURNED TO MEDTRONIC FOR FAILURE INVESTIGATION. THE PUMP WAS OBSERVED TO MAKE A CHATTERING SOUND DURING OPERATION. UPON FURTHER INSPECTION OF THE PUMP, ONE PISTON BUSHING WAS FOUND TO BE DISLODGED FROM THE SMALL CAP END OF THE PUMP HOUSING. DEBRIS WAS OBSERVED NEAR THE LINEAR PISTON BEARINGS, INDICATING WEAR IN THE BEARINGS. NO SIGNS OF WEAR WAS OBSERVED IN THE DOGBONE BEARINGS. NO CORRECTIVE ACTION IS REQUIRED, BECAUSE NO TREND HAS BEEN IDENTIFIED. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT, WHILE IN USE ON A PATIENT, THE HT70 PLUS VENTILATOR GENERATED AN ABNORMAL NOISE. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR. THE PATIENT WAS NOT HARMED OR INJURED AS A RESULT OF THE REPORTED EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
837303 HT70 PLUS VENTILATOR CONTINUOUS VENTILATOR CBK NEWPORT MEDICAL INSTRUMENTS, INC HT70

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention