COBAS 6000 E 601 MODULE
Report
- Report Number
- 1823260-2016-01983
- Event Type
- Malfunction
- Date Received
- December 19, 2016
- Date of Event
- November 29, 2016
- Report Date
- January 13, 2017
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- JJE
- PMA / PMN Number
- K060373
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). THIS EVENT OCCURRED IN (B)(6).
A SPECIFIC ROOT CAUSE COULD NOT BE DETERMINED. THIS TYPE OF SCENARIO USUALLY INDICATES AN ISSUE WITH THE BEAD CAPTURING. THE TYPICAL CAUSE FOR THIS IS CONTAMINATION OF THE LIQUID FLOW PATH OR MEASURING CELL DUE TO A PREANALYTIC ISSUE WITH THE SAMPLE, SUCH AS INSOLUBLE MATERIAL, ANTI-COAGULATION TREATMENT, OR HIGH PROTEIN CONTAINING SAMPLES. THE ANALYZER WAS CHECKED AND FOUND WORKING WITHIN SPECIFICATION AND PERFORMANCE TESTING WAS DONE AND WAS ACCEPTABLE. NO FURTHER ISSUES WERE REPORTED.
THE CUSTOMER EXPERIENCED AN ISSUE WITH QUESTIONABLE RESULTS FOR MULTIPLE ASSAYS ON (B)(6) 2016. ON (B)(6) 2016, THE LABORATORY WAS REQUESTED TO REPEAT ELECSYS FT4 II ASSAY TESTING FOR SOME SAMPLES WITH HIGH RESULTS. AFTER INVESTIGATION, IT WAS FOUND ALL SAMPLES TESTED ON ONE MEASURING CELL OF THE ANALYZER WERE AFFECTED. THE TOTAL NUMBER OF PATIENT SAMPLES INVOLVED WAS UNKNOWN. OF THE DATA PROVIDED, 172 DISCREPANT RESULTS WERE RECEIVED FOR THE ELECSYS FOLATE III ASSAY, 18 DISCREPANT RESULTS WERE RECEIVED FOR THE ELECSYS FT4 II ASSAY, 11 DISCREPANT RESULTS WERE RECEIVED FOR TROPONIN T HS, AND 11 DISCREPANT RESULTS WERE RECEIVED FOR THE ELECSYS PROLACTIN ASSAY. REFER TO THE ATTACHMENT TO THE MEDWATCH FOR ALL PATIENT DATA. NO UNITS OF MEASURE WERE PROVIDED. ALL OF THE ERRONEOUS RESULTS WERE REPORTED OUTSIDE THE LABORATORY. THE PATIENTS WERE NOT ADVERSELY AFFECTED. THE REAGENT LOT NUMBERS WERE FOLATE 144782, FT4 180429, TROPONIN T HS 138696, AND PROLACTIN 143577. THE EXPIRATION DATES WERE REQUESTED BUT WERE NOT PROVIDED. ROUTINE MAINTENANCE WAS PERFORMED ON (B)(6) 2016 AND QC PERFORMED ON THE ANALYZER ON (B)(6) 2016 WAS ACCEPTABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 837690 | COBAS 6000 E 601 MODULE | IMMUNOCHEMISTRY ANALYZER | JJE | ROCHE DIAGNOSTICS | E601 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |