FDA Adverse Event Malfunction Summary report: N

X-SMART ENDO MOTOR

MDR report key: 6182886 · Received December 16, 2016

Report

Report Number
2320721-2016-00067
Event Type
Malfunction
Date Received
December 16, 2016
Date of Event
November 16, 2016
Report Date
February 24, 2017
Manufacturer
DENTSPLY TULSA DENTAL SPECIALTIES
Product Code
EKX
PMA / PMN Number
K990682
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

CONSOLE IS RUNNING TO MANUFACTURER'S SPECIFICATIONS AND FUNCTIONING PROPERLY. MICRO-MOTOR HAS BEEN DAMAGED WITH THE CASING CRACKING ON THE INSIDE. F16R IS OPERATING TO MANUFACTURER'S SPECIFICATIONS.

Additional Manufacturer Narrative · 1

THERE HAS BEEN A PREVIOUS REPORT RECEIVED WHERE THIS MALFUNCTION HAS CAUSED FILE SEPARATION. SINCE SEPARATION OF A FILE COULD NECESSITATE MEDICAL/SURGICAL INTERVENTION TO PRECLUDE PERMANENT DAMAGE TO A BODY STRUCTURE OR PERMANENT IMPAIRMENT OF A BODY FUNCTION, THIS MALFUNCTION WOULD BE LIKELY TO CAUSE/CONTRIBUTE TO A SERIOUS INJURY SHOULD IT RECUR. AS SUCH, THIS EVENT MEETS THE CRITERIA FOR REPORTABILITY PER 21 CFR PART 803. THE DEVICE IS AVAILABLE FOR EVALUATION, THOUGH HAS NOT BEEN RETURNED AS OF THIS REPORT. EVALUATION RESULTS WILL BE SUBMITTED AS THEY BECOME AVAILABLE.

Description of Event or Problem · 1

IN THIS EVENT IT WAS REPORTED THAT AN X-SMART ENDO MOTOR WAS "GOING TOO FAST"; NO INJURY RESULTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
833586 X-SMART ENDO MOTOR HANDPIECE, DIRECT DRIVE, AC-POWERED EKX DENTSPLY TULSA DENTAL SPECIALTIES NA NA

Patients

Seq Age Sex Outcome Treatment
1