FDA Adverse Event Malfunction Summary report: N

LIFECHOICE

MDR report key: 6182777 · Received December 16, 2016

Report

Report Number
3008185181-2016-00001
Event Type
Malfunction
Date Received
December 16, 2016
Date of Event
September 28, 2016
Report Date
December 16, 2016
Manufacturer
INOVA LABS, INC.
Product Code
CAW
PMA / PMN Number
K072688
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

INOVA LABS, INC, THE LEGAL MANUFACTURER OF THE LIFECHOICE PORTABLE OXYGEN CONCENTRATOR, PERSONNEL LEARNED ON 18-NOV-2016 THAT RESMED CORPORATE PERSONNEL SUBMITTED MEDWATCH 3004604967-2016-01167 FOR THIS MALFUNCTION. THE SUBMISSION OF 304604967-2016-01167 WAS PERFORMED IN ERROR BY RESMED PERSONNEL. INOVA LABS, INC. IS SUBMITTING THIS MEDWATCH #3008185181-2016-00001 TO PROVIDE CORRECTED INFORMATION ABOUT THIS EVENT AND THE DEVICE MANUFACTURER. (B)(4). THE ROOT CAUSE OF THE MALFUNCTION WAS DETERMINED TO BE A LOW INTERNAL RESISTANCE (NEAR SHORT) IN THE PIN/BARREL JACK OF THE AC/DC POWER SUPPLY ACCESSORY CORD. WEAR AND/OR ABUSE CONTRIBUTED TO THE CONDITION THAT CAUSED THIS MALFUNCTION. THIS REPORT WAS INITIATED AS THE MALFUNCTION, ALBEIT UNLIKELY GIVEN THE SELF-EXTINGUISHING CHARACTERISTICS OF THE DEVICE MATERIAL, COULD CAUSE OR CONTRIBUTE TO A SERIOUS INJURY IF IT WERE TO RECUR AND PRODUCE A VISIBLE FLAME. IT IS UNKNOWN IN THIS INCIDENT IF THERE WAS AN ACTUAL FLAME, THUS THIS MALFUNCTION IS BEING REPORTED USING CONSERVATIVE REPORTING PRACTICES.

Description of Event or Problem · 1

IT WAS REPORTED THAT A LIFECHOICE OXYGEN CONCENTRATOR POWER SUPPLY CAUGHT FIRE. NO PATIENT INJURY WAS REPORTED AS A RESULT OF THIS EVENT. NO DAMAGE TO PERSON OR PROPERTY WAS REPORTED AS A RESULT OF THIS EVENT. QUOTE OF EVENT AS RECEIVED: "CAUGHT ON FIRE. THE CHARGER PLUGGED IN AND THE WIRE CAUGHT FIRE. NO INJURY, NO DAMAGES OCCURRED."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
830982 LIFECHOICE PORTABLE OXYGEN CONCENTRATOR CAW INOVA LABS, INC. XYC100 N/A

Patients

Seq Age Sex Outcome Treatment
1