FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 6182767
·
Received December 16, 2016
Report
- Report Number
- 3008642652-2016-09302
- Event Type
- Malfunction
- Date Received
- December 16, 2016
- Date of Event
- November 13, 2016
- Report Date
- December 16, 2016
- Manufacturer
- ZOLL MANUFACTURING CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (SERVICE CODE 204) HAS BEEN CONFIRMED. AS RECEIVED, THE ELECTRODE BELT FAILED INCOMING FUNCTIONALITY TESTING. THE CAUSE FOR THE SERVICE CODE 204 AND TEST FAILURE WAS ISOLATED TO AN OPEN YELLOW (PGND) WIRE IN THE TRUNK CABLE. THE ROOT CAUSE FOR THE OPEN WIRE WAS EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE ELECTRODE BELT.
Description of Event or Problem · 1
A US DISTRIBUTOR RETURNED ELECTRODE BELT SN (B)(4) AND REPORTED THAT A PATIENT HAD RECEIVED A SERVICE CODE 204; "BELT/MONITOR UNUSABLE".