AMIS LEG POSITIONER
Report
- Report Number
- 3005180920-2016-00669
- Event Type
- Injury
- Date Received
- December 16, 2016
- Date of Event
- November 18, 2016
- Report Date
- March 17, 2017
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- FWX
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
ON 13 MARCH 2017 THE R&D PROJECT MANAGER PERFORMED A VISUAL INSPECTION OF THE RETRIEVED ITEM AND COMMENTED AS FOLLOWS: FOLLOWING VISUAL AND FUNCTIONAL CHECKS OF THE ITEM, NO IMPROPRIERTIES WERE FOUND.
ADDITIONAL INFORMATION RECEIVED FROM THE INITIAL REPORTER ON (B)(6) 2016 AND INCLUDES: THE NURSE WAS « EXPERT » AND USED TO MANIPULATE THE LEG POSITIONER BUT FIRST THE ROTATION WAS UNEASY AND THEN PROBABLY HAPPENED THE FRACTURE. HOWEVER THEY FINALLY NOTICED THE FRACTURE AT THE END OF THE INTERVENTION DURING THE REDUCTION. THE FRACTURE WAS FIXED DURING THE SAME INTERVENTION. ON (B)(6) 2016 THE MEDICAL AFFAIRS DIRECTOR PERFORMED A CLINICAL EVALUATION AND COMMENTED AS FOLLOWS: THE SUCCESSION OF EVENTS IS NOT VERY CLEAR: APPARENTLY, A TIBIAL FRACTURE WAS GENERATED DURING THE THA OPERATION IN A VERY OSTEOPOROTIC PATIENT (ACCORDING TO REPORT), POSSIBLY DURING THE EXTERNAL ROTATION OF THE LIMB. THIS OPERATION IS DONE MANUALLY BY AN OR HELP, NORMALLY. IT IS NOT CLEAR HOW AND WHY MEDACTA DEVICES WOULD BE RESPONSIBLE FOR THIS EVENT. NO FINAL CONCLUSION CAN BE DRAWN WITH THE INFORMATION CURRENTLY AVAILABLE. BATCH REVIEWS PERFORMED ON (B)(6) 2016. AMIS MOBILE LEG POSITIONER 2.0, CODE 01.15.10.0190, LOT. 1411958 ONE ITEM MANUFACTURED AND RELEASED ON (B)(6) 2014. NO ANOMALIES FOUND RELATED TO THE PROBLEM. AMIS SHOE, CODE 01.15.10.0315 LOT 1551837: THE 30 ITEMS MANUFACTURED AND RELEASED ON (B)(6) 2015. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, NO SIMILAR EVENTS HAVE BEEN REPORTED ON ITEMS OF THE SAME LOT.
AT THE END OF THE HIP SURGERY IT WAS NOTICED THAT THE TIBIA AROUND MALLEOLUS OF THE PATIENT WAS FRACTURED. THE NURSES IN THE OR FIRST HAD A HARD MANEUVER WITH EXTERNAL ROTATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 831175 | AMIS LEG POSITIONER | TABLE FOR SURGERY | FWX | MEDACTA INTERNATIONAL SA | 1411958 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |