FDA Adverse Event Malfunction Summary report: N

CENTRALINK DATA MANAGEMENT SYSTEM

MDR report key: 6179544 · Received December 15, 2016

Report

Report Number
2432235-2016-00770
Event Type
Malfunction
Date Received
December 15, 2016
Date of Event
November 21, 2016
Report Date
December 15, 2016
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS INC.
Product Code
JQP
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A SIEMENS CUSTOMER SERVICE ENGINEER (CSE) SPECIALIST WAS DISPATCHED TO THE CUSTOMER'S SITE. THE CSE REMOVED UNNECESSARY FILES FROM THE DRIVE AND RESTORED A DATABASE BACKUP. COMMUNICATION ISSUES WERE STILL OCCURRING. FURTHER EVALUATION BY THE CSE FOUND THE PROBLEM WAS RELATED TO THE NETWORK CARD. THE CSE REPLACED THE NETWORK CARD. THE CSE ALSO REPLACED THE "BACKUP" SERVER WITH ANOTHER SERVER THAT HAS MORE SPACE. THE CAUSE OF THE DELAY IN UPLOADING PATIENT RESULTS WAS DUE TO THE NETWORK CARD.THE INSTRUMENT IS PERFORMING ACCORDING TO SPECIFICATIONS.NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.

Description of Event or Problem · 1

CUSTOMER REPORTED THEIR LOCAL "D" DRIVE WAS FULL, CAUSING A DELAY IN UPLOADING PATIENT SAMPLE RESULTS FROM THEIR CENTRALINK DATA MANAGEMENT SYSTEM. THE CUSTOMER ALSO STATED THERE WAS NO COMMUNICATION BETWEEN CENTRALINK AND THEIR LABORATORY INFORMATION SYSTEM (LIS), DELAYING 1,000 REQUESTS, LEAVING THEM IN THE VALIDATED STATUS. THERE ARE NO REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DELAY IN UPLOADING PATIENT SAMPLE RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
827336 CENTRALINK DATA MANAGEMENT SYSTEM DATA MANAGER JQP SIEMENS HEALTHCARE DIAGNOSTICS INC. CENTRALINK DATA MANAGEMENT SYSTEM

Patients

Seq Age Sex Outcome Treatment
1