FDA Adverse Event
Injury
Summary report: N
INFINIA II GANTRY
MDR report key: 617874
·
Received June 28, 2005
Report
- Report Number
- 9613299-2005-00002
- Event Type
- Injury
- Date Received
- June 28, 2005
- Date of Event
- June 7, 2005
- Report Date
- June 28, 2005
- Manufacturer
- GE MEDICAL SYSTEMS F.I. HAIFA
- Product Code
- KPS
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
IT WAS REPORTED THAT A PT'S WRIST WAS BROKEN WHILE UNDER GOING A SCAN USING THE ININIA II. AS PART OF THE STATIC SCAN THE HCP REQUESTED THE PT TO PLACE THEIR OUTSTRETCHED. LEFT ARM ON A TABLE TRAY WHICH HAD BEEN POSITIONED NEXT TO GANTRY IN A POSTED, NO OBSTACLE ZONE. THE SCAN WAS UNDERTAKEN WITHOUT INCIDENT. UPON COMPLETION, THE HCP INITIATED GANTRY MOTION IN ORDER TO REMOVE THE PT. DURING ROTATIONAL MOTION, HEAD #1 COLLIDED WITH THE TRAY TABLE CAUSING THE INJURY TO THE PT'S WRIST. THE PSD STOPPED FURTHER ROTATION. THE PT WAS TREATED AT THE HOSP'S EMERGENCY DEPARTMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INFINIA II GANTRY | NUCLEAR MEDICINE | KPS | GE MEDICAL SYSTEMS F.I. HAIFA | ASM001465 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 85 YR | Required Intervention |