E SERIES DEFIBRILLATOR
Report
- Report Number
- 1220908-2016-03121
- Event Type
- Malfunction
- Date Received
- December 15, 2016
- Report Date
- November 28, 2016
- Manufacturer
- ZOLL MEDICAL CORPORATION
- Product Code
- MKJ
- PMA / PMN Number
- K042007
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
ZOLL MEDICAL CORPORATION HAS RECEIVED THE PRODUCT AND WILL BE PROVIDING A FOLLOW-UP REPORT WHEN OUR INVESTIGATION IS COMPLETED.
THE MALFUNCTION WAS OBSERVED DURING INCOMING TESTING; HOWEVER AFTER DISASSEMBLING THE DEVICE TO DETERMINE ROOT CAUSE, THE MALFUNCTION WAS NO LONGER SEEN. THE DEVICE WAS PUT THROUGH EXTENSIVE TESTING WITHOUT DUPLICATING THE MALFUNCTION. A FLEX CABLE WAS REPLACED AS A PRECAUTION. THE DEVICE PASSED THE FINAL TEST PROCEDURE, WAS RECERTIFIED, AND RETURNED TO THE CUSTOMER. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE FAILED SELF TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 827917 | E SERIES DEFIBRILLATOR | E SERIES | MKJ | ZOLL MEDICAL CORPORATION | E SERIES | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |