FDA Adverse Event Malfunction Summary report: N

E SERIES DEFIBRILLATOR

MDR report key: 6178033 · Received December 15, 2016

Report

Report Number
1220908-2016-03121
Event Type
Malfunction
Date Received
December 15, 2016
Report Date
November 28, 2016
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
PMA / PMN Number
K042007
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ZOLL MEDICAL CORPORATION HAS RECEIVED THE PRODUCT AND WILL BE PROVIDING A FOLLOW-UP REPORT WHEN OUR INVESTIGATION IS COMPLETED.

Additional Manufacturer Narrative · 1

THE MALFUNCTION WAS OBSERVED DURING INCOMING TESTING; HOWEVER AFTER DISASSEMBLING THE DEVICE TO DETERMINE ROOT CAUSE, THE MALFUNCTION WAS NO LONGER SEEN. THE DEVICE WAS PUT THROUGH EXTENSIVE TESTING WITHOUT DUPLICATING THE MALFUNCTION. A FLEX CABLE WAS REPLACED AS A PRECAUTION. THE DEVICE PASSED THE FINAL TEST PROCEDURE, WAS RECERTIFIED, AND RETURNED TO THE CUSTOMER. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE FAILED SELF TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
827917 E SERIES DEFIBRILLATOR E SERIES MKJ ZOLL MEDICAL CORPORATION E SERIES NA

Patients

Seq Age Sex Outcome Treatment
1