FDA Adverse Event Injury Summary report: N

ARTICULEZE M HEAD 36MM +5

MDR report key: 6176642 · Received December 15, 2016

Report

Report Number
1818910-2016-33515
Event Type
Injury
Date Received
December 15, 2016
Date of Event
July 28, 2016
Report Date
December 13, 2016
Manufacturer
DEPUY ORTHOPAEDICS INC US
Product Code
JDI
PMA / PMN Number
K980513
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
ATTORNEY

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISION OF 21 CFR, PART 803. THE REPORT MAY BE BASED ON THE INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. PRODUCT COMPLAINT #: (B)(4). INVESTIGATION SUMMARY: NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. BASED ON PREVIOUS INVESTIGATIONS THIS COMPLICATION OF JOINT REPLACEMENT IS UNLIKELY TO HAVE BEEN THE RESULT OF A DEVICE FAILING TO MEET REQUIRED SPECIFICATIONS. THE INFORMATION RECEIVED WILL BE RETAINED FOR POTENTIAL SERIES INVESTIGATIONS IF TRIGGERED BY TREND ANALYSIS, POST MARKET SURVEILLANCE, OR OTHER EVENTS WITHIN THE QUALITY SYSTEM. CORRECTIVE ACTION WAS NOT INDICATED. DEVICE HISTORY LOT: NULL. DEVICE HISTORY BATCH: NULL. DEVICE HISTORY REVIEW: NULL. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. A WORLDWIDE LOT SPECIFIC COMPLAINT DATABASE SEARCH, OR DEVICE HISTORY RECORD (DHR) REVIEW WAS NOT POSSIBLE BECAUSE THE REQUIRED LOT CODE(S) WAS NOT PROVIDED. BASED ON PREVIOUS INVESTIGATIONS THIS COMPLICATION OF JOINT REPLACEMENT IS UNLIKELY TO HAVE BEEN THE RESULT OF A DEVICE FAILING TO MEET REQUIRED SPECIFICATIONS. THE INFORMATION RECEIVED WILL BE RETAINED FOR POTENTIAL SERIES INVESTIGATIONS IF TRIGGERED BY TREND ANALYSIS, POST MARKET SURVEILLANCE, OR OTHER EVENTS WITHIN THE QUALITY SYSTEM. CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. UDI: UNAVAILABLE. THIS COMPLAINT IS THE SUBJECT OF LITIGATION OR A LEGAL CLAIM AND CURRENTLY COMPLETE PRODUCT DETAIL IS NOT AVAILABLE AT THIS TIME. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

THIS COMPLAINT IS PART OF A LITIGATION CASE, INVOLVING THE PATIENT'S OTHER HIP (LEFT). PATIENT'S RIGHT HIP REVISED TO ADDRESS SIGNIFICANT PAIN. REVISING SURGEON INDICATED THAT THE RIGHT HIP WAS IMPLANTED ELSEWHERE, NO DATE GIVEN; PATIENT HAD SIGNIFICANT PAIN, AND METAL ION LEVELS WERE ELEVATED. INTRAOPERATIVELY, SURGEON FOUND "SIGNIFICANT CORROSION" ON THE TRUNNION. METAL ION LEVELS PRESENT < 7.0 PPB NOT SUPPORTING METAL ION TOXICITY. CORROSION:TAPER ADDED TO STEM AND HEAD PRODUCT HARMS.

Description of Event or Problem · 1

IN ADDITION TO WHAT WERE PREVIOUSLY ALLEGED, PPF ALLEGES METAL WEAR, METALLOSIS AND ELEVATED METAL IONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
829680 ARTICULEZE M HEAD 36MM +5 ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS JDI DEPUY ORTHOPAEDICS INC US 2621036

Patients

Seq Age Sex Outcome Treatment
1 60 YR Required Intervention